Section 7: TSP command reference
Models 707B and 708B Switching Matrix Reference Manual
7-58
707B-901-01 Rev. B / January 2015
Example
digio.trigger[4].pulsewidth = 20e-6
Sets the pulse width for trigger line 4 to
20
μs.
Also see
(on page 7-53)
(on page 7-58)
(on page 7-58)
digio.trigger[N].release()
This functionreleases an indefinite length or latched trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.trigger[
N
].release()
N
Digital I/O trigger line (1 to 14)
Details
Releases a trigger that was asserted with an indefinite pulsewidth time. It also releases a trigger that
was latched in response to receiving a synchronous mode trigger. Only the specified trigger line is
affected.
Example
digio.trigger[4].release()
Releases digital I/O trigger line 4.
Also see
(on page 7-53)
digio.trigger[N].reset()
This function resets trigger values to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.trigger[
N
].reset()
N
Digital I/O trigger line (1 to 14)