Section 3: Connections
Model 6517B Electrometer User's Manual
3-10
6517B-900-01 Rev. B August 2022
Figure 19: Safety shield
1
Device or circuit under test
2
Metal safety shield
3
Noise or guard shield
4
Connect to 6517B guard using a triaxial cable
Safety earth ground
Floating circuits
Many measurements are performed above earth ground and, in some test situations, can result in
safety concerns. The following figures show examples where the 6517B floats at a hazardous
voltage level.
In the following figure, a shock hazard (100 V) exists between meter input LO and chassis ground. If
meter input LO is connected to a noise shield, the shock hazard is also present on that shield.