In this section:
Basic source-measure connections ......................................... 2-1
Test fixture and device under test (DUT) connections ........... 2-20
Basic source-measure connections
This section describes basic information on connecting source-measure units (SMUs), the
preamplifier, and the ground unit to devices under test (DUTs).
The 4200A-SCS is provided with an interlock circuit that must be positively activated for the
high voltage output to be enabled. The interlock helps facilitate safe operation of the
equipment in a test system. Bypassing the interlock could expose the operator to hazardous
voltages that could result in personal injury or death.
Asserting the interlock allows the SMU and preamplifier terminals to become hazardous,
exposing the user to possible electrical shock that could result in personal injury or death.
SMU and preamplifier terminals should be considered hazardous even if the outputs are
programmed to be low voltage. Precautions must be taken to prevent a shock hazard by
surrounding the test device and any unprotected leads (wiring) with double insulation that is
rated for 250 V, Category O.
The maximum allowed voltage between circuit COMMON and chassis ground is ±32 V dc. The
maximum allowed voltages between the preamplifier signals are:
COMMON to chassis ground: 32 V
PEAK
GUARD to COMMON: 250 V
PEAK
SENSE or FORCE to GUARD: 40 V
PEAK
Section 2
Connections and configuration