Proper connection of the supply to the DUT 3
Figure 2
Rear panel view of Model 2306
Figure 3
4-wire remote sense connection of the DUT to the output
NOTE
DO NOT jumper the sense inputs and supply outputs at the rear of the supply.
DO NOT pass the source and sense leads together in the same twisted pair.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ 150VA MAX.
LINE FUSE
SLOWBLOW
2.0A, 250V
IEEE-488
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
+
+
+
_
_
_
_
+
SOURCE
SENSE
DVM
IN
SOURCE
OUTPUT #1
ISOLATION FROM EARTH: 22 VOLTS MAX.
RELAY
CONTROL
24VDC MAX.
+
+
+
_
_
_
_
+
SOURCE
SENSE
DVM
IN
SOURCE
OUTPUT #2
DVM IN
+30 VDC MAX.
CAT
I
External Test Circuit
Quick Disconnect
Connector
Model 2306
source input/output
DVM input
DUT
+
-
DVM +
DVM -
Source -
Source -
Sense -
Sense +
Twisted pair
Figure 2 shows a rear panel view of the Model 2306 detailing the connector sockets for the
battery channel “OUTPUT #1” and the charger channel “OUTPUT #2.” The Model 2306 does
not have an internal local sense connection. Therefore, the operator must connect the DUT to
the supply in either a remote or local sense configuration as illustrated. For a 4-wire remote sense
application, the sense inputs to the supply must be connected at load as close to the inputs of the
load as possible through twisted pair leads as shown in Figure 3.
On some sensitive or easily damaged devices under test (DUTs), the
instrument power-down sequence can apply transient signals to the DUT that
may affect or damage it. When testing this type of DUT, disconnect it from
the instrument before turning the instrument off.
NOTE