Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-43
:BATTery[n]:SIMulator:METHod
This command sets the simulation mode in the battery simulation function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
*RST
Save settings
DYNamic
Usage
:BATTery[n]:SIMulator:METHod <mode>
:BATTery[n]:SIMulator:METHod?
[n]
Channel number: For single-channel instruments, the value is 1
<mode>
DYNamic: Output depends on the running of the battery model
STATic: Constant output
Details
This command sets the battery simulation mode.
If you choose Dynamic, the Voc and SOC of the simulated battery change based on the charging and
discharging, like a real battery.
If you choose Static, the Voc and SOC of the simulated battery stay the same.
Example
:BATT:SIM:METH STAT
Set the simulation mode to static mode.
Also see
(on page 7-7)
(on page 2-110)