Section 3: Functions and features
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
3-48
077114601 / March 2019
List triggering
You must configure list triggering before running a list. List triggering is similar to common triggering.
However, list triggering introduces dwell time to control the source-delay-measure cycle. For more
information about the common triggering, refer to
(on page 3-30).
When running a list from the front panel, you only need to configure the arm source and a sample
count for the trigger model. For other parameters, the instrument changes the settings according to
the list configuration.
Parameters
Value
Trigger source
Immediate
Trigger count
List point
Arm count
List cycle
Initiate continuous state
Off
Once the instrument finishes the list, it goes to idle mode. You can press the
TRIGGER
key to run
the list again.
When executing a list using SCPI commands, you need to set up the trigger model to match the list
configuration. For details, refer to
Configure and execute a 10-step linear list sweep
(on page 5-3).