Master-Slave Series Overview
Background
When connecting 2260B power supplies in
series, up to 2 units can be used in series and
all units must be of the same model.
When the units are used in series, a number of
precautions and limitations apply. Please read
this overview before operating the power
supplies in series.
Master
Load
Ext-V
Ext-R
Slave
VMON
IMON
Limitations
Display
Only the master unit will display the current.
Master and slave units display the voltage. The
total voltage is the sum of the units.
OVP/OCP
The master unit can shut down the slave unit
when OVP/OCP is tripped on the master unit
(if the slave connector is wired for shut down
on alarm).
OVP and OCP level is determined by the
master OVP and OCP level. The OVP and OCP
level on the slave unit is ignored.