background image

Artisan Technology Group 

is your source for quality 

new and certified-used/pre-owned equipment

• 

FAST SHIPPING AND 

DELIVERY

• 

TENS OF THOUSANDS OF 

IN-STOCK ITEMS

•  EQUIPMENT DEMOS
• 

HUNDREDS OF 

MANUFACTURERS 

SUPPORTED

•  LEASING/MONTHLY

 

RENTALS

•  ITAR CERTIFIED

           

SECURE ASSET 

SOLUTIONS

SERVICE CENTER REPAIRS

Experienced engineers and technicians on staff 

at our full-service, in-house repair center

WE BUY USED EQUIPMENT

Sell your excess, underutilized, and idle used equipment 

We also offer credit for buy-backs and trade-ins

www.artisantg.com/WeBuyEquipment

                        

REMOTE INSPECTION

Remotely inspect equipment before purchasing with 

our interactive website at

 

www.instraview.com

LOOKING FOR MORE INFORMATION? 

Visit us on the web at

 

www.artisantg.com

    

for more 

information on price quotations, drivers, technical 

specifications, manuals, and documentation

Contact us:

 

 

(888) 88-SOURCE  |  [email protected]  |  www.artisantg.com

SM

View

Instra

Содержание IntelliFrame VX1410A

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...icing instructions are for use by qualified personnel only To avoid personal injury do not perform any servicing unless you are qualified to do so Refer to the Safety Summary prior to performing servi...

Страница 3: ...all previously published material Specifications and price change privileges reserved Printed in the U S A Tektronix Inc P O Box 1000 Wilsonville OR 97070 1000 TEKTRONIX TEK and IntelliFrame are regi...

Страница 4: ...fect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be obligated to furnish service under this warranty a to repair damage resulting from a...

Страница 5: ...rtification on delivery plus two more annual calibrations from your ser vice center Five years of calibration services Option C5 Provides initial Certification on delivery plus four more annual calibr...

Страница 6: ...or Connector 3 1 Enhanced Monitor RS 232 Connector 3 4 Backplane Connectors 3 5 Enhanced Monitor 3 9 Command Groups 3 13 Display Commands 3 13 Status Commands 3 13 System Commands 3 16 Test Commands 3...

Страница 7: ...erification Equipment Required 5 1 DC Voltage Regulation 5 1 DC Load Ripple and Noise 5 2 Theory of Operation IntelliFrame Mainframe 6 1 Mainframe with Option 1M 6 3 Adjustment Procedures Adjustment P...

Страница 8: ...ses 8 21 Fault Isolation 8 21 Replaceable Parts Parts Ordering Information 9 1 IntelliFrameR mainframe Common Replaceable Parts 9 2 Using the Replaceable Parts List 9 3 Appendices Appendix A Power Bud...

Страница 9: ...panel fillers 1 21 Figure 1 13 Locations of safety and chassis ground connections 1 22 Figure 1 14 Installing the EMI intermodule shield and front panel EMI fillers 1 23 Figure 1 15 Logical address sw...

Страница 10: ...ront view and side view of IntelliFrame Mainframe with Option 2U and Rackmount Option 4 13 Figure 6 1 Mainframe block diagram 6 2 Figure 8 1 Removing the fan blower filters 8 3 Figure 8 2 Removing the...

Страница 11: ...mbly 9 13 Figure 9 5 1R Rackmount assembly 9 15 Figure 9 6 2R Rackmount and door assembly 9 17 Figure 9 7 Option 1U cable tray 9 19 Figure 9 8 Option 2U cable tray 9 21 Figure 9 9 Display panel VX1420...

Страница 12: ...ds 3 19 Table 3 12 IEEE 488 2 common commands 3 20 Table 3 13 Parameter types for syntax descriptions 3 23 Table 3 14 Default factory RS 232 communication settings 3 111 Table 3 15 Effects of PRESet o...

Страница 13: ...232 settings for the calibration Procedure 7 2 Table 8 1 IntelliFrame Mainframe Enhanced Monitor jumper configurations 8 11 Table 8 2 Fuses 8 21 Table 9 1 IntelliFrame Mainframe common replaceable par...

Страница 14: ...e or shock hazard observe all ratings and markings on the product Consult the product manual for further ratings information before making connections to the product The common terminal is at ground p...

Страница 15: ...ty Terms on the Product These terms may appear on the product DANGER indicates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately access...

Страница 16: ...of rendering first aid and resuscitation is present Disconnect Power To avoid electric shock disconnect the main power by means of the power cord or if provided the power switch Use Care When Servicin...

Страница 17: ...Service Safety Summary xii VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 18: ...ion for the Enhanced Monitor option H Performance Verification includes the performance verification procedures used to verify the advertised specifications of the mainframe H Theory of Operation incl...

Страница 19: ...com For product support outside of North America contact your local Tektronix distributor or sales office Service Support Contact your local Tektronix distributor or sales office Or visit our web site...

Страница 20: ...ations worldwide or you can visit us on our Customer Services World Center web site at http www tek com Mea surement Service See our latest service offerings and contact us by email In addition to war...

Страница 21: ...ix supports repair to the replaceable part level by providing for circuit board exchange Use this service to reduce down time for repair by exchanging circuit boards for remanufactured ones Tektronix...

Страница 22: ...Getting Started Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 24: ...nal for VX1410A H EMI DIN shields for superior EMI protection H Optional adjustable rackmounts and cable trays These features are explained in more detail in the following paragraphs The IntelliFrame...

Страница 25: ...ith higher cooling capacity and very quiet operation The higher cooling fan operates at a low fan speed with equivalent airflow The IntelliFrame Mainframe has an auto configuration solid state backpla...

Страница 26: ...y via the VXIbus or a rear RS 232 interface The VX1410A Option 1M includes a 20 character alphanumeric display on the front panel with scroll keys for manual viewing for mainframe status The VX1420A c...

Страница 27: ...varies according to country see Table 1 2 H Fuse holder H Line fuse 15 A 250 V H Line fuse 20 A 250 V H Line fuse 6 3 A 250 V H VXIplug play Floppy disk See the Readme file on disk for installation i...

Страница 28: ...r Tektronix part number 119 5061 XX H US Power cord for the external supply Tektronix part number 161 0213 00 H Option A1 Power cord for the external supply Tektronix part number 161 0209 00 H Option...

Страница 29: ...ber 161 0218 XX For high mainframe loads combined with low input line voltages shaded region use only the power cord with the 20 A plug Select the proper fuse based on the ranges shown in Figure 1 1 M...

Страница 30: ...orx and Phillips driver to install Options 1U or 2U Options 1U and 2U can be installed three different ways without rails with rails or with rails installed at 90_ Figure 1 2 on page 1 8 and Figure 1...

Страница 31: ...l Opt 1U without rails 1 75 inch depth Install 4 screws and leave 1 8 inch clearance from the chassis Slide the cable tray over the screws then tighten screws to secure cable tray Figure 1 2 Install O...

Страница 32: ...to secure rails Install 4 screws and leave 1 8 inch clearance from the rails Slide the cable tray over the screws then tighten screws to secure cable tray 1 2 2 Install 4 screws and leave 1 8 inch cl...

Страница 33: ...nal Opt 2U without rails 3 5 inch depth Remove feet 4 Install 4 screws and leave 1 8 inch clearance from the chassis Slide the cable tray over the screws then tighten screws to secure cable tray Figur...

Страница 34: ...s to secure rails 1 Install 4 screws and leave 1 8 inch clearance from the rails Slide the cable tray over the screws then tighten screws to secure cable tray 2 Install 4 screws and leave 1 8 inch cle...

Страница 35: ...gnal from the front panel This feature allows you to reset the complete VXI system without having to power down the mainframe A SYSReset activation is accomplished by pushing the right side nonlighted...

Страница 36: ...ss the mainframe in the rack position the rackmount bracket as shown on the left side of Figure 1 6 2 Determine how far you want to recess or extend the IntelliFrame Mainframe and install the rackmoun...

Страница 37: ...ceable Parts on page 9 15 Orient the button latch so that it faces away from the mainframe and toward the rear Use the following procedure to install the IntelliFrame Mainframe in the rack 1 Identify...

Страница 38: ...gure 1 7 Leave the screws loose so that you can adjust the overall length of the slide out track assembly when installing the mainframe in the rack Use the extended length brackets to comply with deep...

Страница 39: ...position of securing holes Figure 1 8 Vertical clearance for rack installation standard mainframe 7 Select the mounting method according to the rack type To mount the slide out tracks with the front...

Страница 40: ...ws 4 Left front rail Figure 1 9 Installing the slide out track assemblies in the rack Top view WARNING Because of the size and weight of the mainframe installing the mainframe alone can result in pers...

Страница 41: ...e 10 32 screws using 28 inch lbs of torque 18 Install the proper power cord at the rear of the mainframe NOTE To ensure the mainframe is properly mounted in the rack do not tighten the screws while in...

Страница 42: ...3 Install the catch on the right rackmount bracket You can adjust the catch position forward or back to assure a flush fit between the door and the rackmount bracket 4 Lift the rackmount door and mate...

Страница 43: ...ire H Investigate the possibility of fire hazards due to the availability of power from the mainframe exceeding 150 W per slot Use similar considerations for VXIbus Modules that externally receive pow...

Страница 44: ...ON To avoid damage caused by heat use only Tektronix front panels otherwise the shutters may activate effectively robbing airflow from installed modules Installing the fillers provides improved coolin...

Страница 45: ...in place to ensure the proper bonding of the power supply to the mainframe You can use the chassis ground connection to connect the grounds of one or more instruments to the mainframe This ensures a...

Страница 46: ...oves of the card guide on one side of each installed VXIbus module This is recommended to use with unshielded modules the shielding may not be compatible with some shielded modules The shield conforms...

Страница 47: ...s However make sure that no other module has the same slot address as the Enhanced Monitor board If all VXI modules in the mainframe are set to FF the resource manager will dynamically configure the m...

Страница 48: ...your mainframe does not include the optional display you can measure the power supply voltages with a digital voltmeter Connect one lead of the voltmeter to the supply pins of the 25 pin connector on...

Страница 49: ...Getting Started 1 26 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 50: ...Operating Basics Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 51: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 52: ...bottom of the mainframe Slot 0 is reserved for the mainframe controller all other slots are available for any VXIbus modules IntelliGuides on the bottom of each slot automatically open when you insta...

Страница 53: ...a 20 character alphanumeric display indicating the status of the system The NEXT and PREV push buttons located to the right of the display allow you to scroll through the display selections See Figure...

Страница 54: ...re 2 3 Rear view of a IntelliFrame Mainframe WARNING To prevent damage to the instrument do not connect an RS 232 cable to the 25 pin connector The connector is not an RS 232 connector Connecting an R...

Страница 55: ...l error messages have been retrieved the READY indicator stops flashing and remains on The amber ACCESSED indicator lights when the MODID line is accessed by the slot 0 device Refer to Enhanced Monito...

Страница 56: ...Reference Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 57: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 58: ...he 25 pin Sub D connector lets you monitor the power supply voltages fan speed and the maximum slot temperature rise within the mainframe The connector also provides remote on and off capability and a...

Страница 59: ...portional to the maximum temperature rise of the 13 modules 100 mV C 0V 0 C 12 Tach1 VX1410A A square wave output signal proportional to the speed of Fan 1 RPM 15 Pulse Width where Pulse Width is the...

Страница 60: ...pins 8 and 21 5 VB to supply 5 V Standby to the backplane make sure that the rear panel jumper is in the correct position refer to Enhanced Monitor Board Jumpers on page B 4 4 Refer to VMEbus Specifi...

Страница 61: ...onitor includes a 9 pin RS 232 connector that allows connec tion to a RS 232 host Figure 3 2 shows the pinouts of the 9 pin RS 232 connector Table 3 2 describes the pin assignments Pin 9 Pin 1 Figure...

Страница 62: ...dy DTR 5 Signal Ground GND 6 No Connection 7 Request to Send RTS 8 Clear to Send CTS 9 No Connection Backplane Connectors Table 3 3 shows the P1 connector pinouts for all slots in the IntelliFrame Mai...

Страница 63: ...BG3IN SYSFAIL 11 GND BG3OUT BERR 12 DS1 BR0 SYSRESET 13 DS0 BR1 LWORD 14 WRITE BR2 AM5 15 GND BR3 A23 16 DTACK AM0 A22 17 GND AM1 A21 18 AS AM2 A20 19 GND AM3 A19 20 IACK GND A18 21 IACKIN SERCLK A17...

Страница 64: ...1 LBUSA04 A31 LBUSC04 12 LBUSA05 GND LBUSC05 13 5 2 V 5 V 2 V 14 LBUSA06 D16 LBUSC06 15 LBUSA07 D17 LBUSC07 16 GND D18 GND 17 LBUSA08 D19 LBUSC08 18 LBUSA09 D20 LBUSC09 19 5 2 V D21 5 2 V 20 LBUSA10 D...

Страница 65: ...ODID08 A31 LBUSC04 12 MODID07 GND LBUSC05 13 5 2 V 5 V 2 V 14 MODID06 D16 LBUSC06 15 MODID05 D17 LBUSC07 16 GND D18 GND 17 MODID04 D19 LBUSC08 18 MODID03 D20 LBUSC09 19 5 2 V D21 5 2 V 20 MODID02 D22...

Страница 66: ...ont panel display is standard includes a front panel display to indicate the system status as determined by the monitoring system Use the NEXT and PREV buttons on the side of the display to scroll thr...

Страница 67: ...IL S2 XX C FAIL S3 XX C FAIL S4 XX C FAIL S5 XX C FAIL S6 XX C FAIL S7 XX C FAIL S8 XX C FAIL S9 XX C FAIL S10 XX C FAIL S11 XX C FAIL S12 XX C FAIL Ambient XX C FAIL Fan1 XXXXRPM FAIL Fan2 XXXXRPM VX...

Страница 68: ...HHHH MM Hr of Oper HHHHHH Date and Time Message Year Mon XX XX XX2 SYSTEM OFF SYSTEM OK 1 To reset the filter cleaning interval display the filter status message by pressing the Next or Prev scroll bu...

Страница 69: ...mation H A summary of the SCPI commands and queries H An overview of the syntax used with the commands and queries H A detailed description of each command and query H A list of error messages H Brief...

Страница 70: ...begins on Page 3 27 Table 3 7 Display commands Header Description Sets or queries user defined messages to display on the front panel readout Sets or queries the marquee status to scroll the display S...

Страница 71: ...Condition register TRIGger Returns the trigger location ENABle Sets or Queries the Contents of the VXITrigger enable register PRESet Presets the SCPI enable registers transition filters and the Error...

Страница 72: ...l Returns the wattage used by the mainframe LIMit Set or query the maximum amount of power allowed before generating an alarm MAXimum Returns the maximum power high mark used since the last factory ma...

Страница 73: ...codes from the Error Event queue Returns the number of unread events from the Error Event queue Sets or queries enable events to be placed in the Error Event queue Presets the Error Event queue enable...

Страница 74: ...RAW Configures the RS 232 port parameters for use with a computer TERMinal Configures the RS 232 port parameters for use with a terminal SBITs Sets or queries the number of stop bits sent with each c...

Страница 75: ...mainframe on or off The query returns whether the power is on or off Queries the number of times the mainframe power has been cycled Queries the power source of the enhanced monitor Sets or queries wh...

Страница 76: ...urns the failure code for the last self test command that was executed Returns a failure code as a string describing the last executed self test command and the test results VXI Monitor Commands Comma...

Страница 77: ...nt Status Enable register Returns the contents of the Standard Event Status Register Returns the identification message of the mainframe Synchronizes command execution with the controller Sets or quer...

Страница 78: ...struments The standard simplifies the task of programming a group of instruments that use SCPI Instead of having to learn different commands for every instrument the programmer may use the same comman...

Страница 79: ...evel nodes Root node Leaf nodes Figure 3 3 Tree hierarchy The header path and leaf node together form the command header Figure 3 4 shows the command header for the leaf node indicated in Figure 3 3 R...

Страница 80: ...al numeric Integers in hexadecimal H octal Q or binary B format command parameter H00FF B010101 Q0753 NR1 Integers response parameter 0 1 15 NR2 Decimal numbers response parameter 1 2 3 141516 6 5 NRf...

Страница 81: ...mand set easier to remember and use The mainframe is not sensitive to the case of command characters You can enter commands in either upper or lower case You can precede any command with white space c...

Страница 82: ...different header path from the root level you must use a colon to force a return to the root level SYSTem VERSion TEST RESult If the command that follows the semicolon has the same header path you may...

Страница 83: ...Syntax 3 26 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 84: ...410A mainframes with Option 1M or field upgrade VXF1M Use the DISPlay subsystem to display user defined messages on the front panel Figure 3 6 lists the DISPlay subsystem parameters NOTE The standard...

Страница 85: ...there are no monitor failures Parameter Query response String Block String Not Applicable 223 Too Much Data You attempted to send too much data to the display 220 Parameter Error You attempted to send...

Страница 86: ...otate to the left regardless of the length DISPlay WINDow TEXT MARQuee Display Control DISPlay WINDow TEXT MARQuee Display control Query response AUTO ON AUTO ON Not Applicable 224 Illegal Parameter V...

Страница 87: ...DISPlay Subsystem 3 30 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 88: ...o until an event occurs If the triggering event disappears the corresponding event bit remains set until the register is read or RST IEEE command is performed H Reading the Condition Register is not d...

Страница 89: ...ame registers since the event still exists the Condition Register will display the problem with 5V current The Event Register was cleared from the previous read operation STAT QUES CURR The system res...

Страница 90: ...no command associated with this query Parameter Query response Not applicable NR1 Operation status word bits Definition D0 CALibrating D1 Not Used D2 Not Used D3 Not Used D4 Not Used D5 VXI Triggers D...

Страница 91: ...4 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Query Response STATus OPERation CONDition Examples Related Commands Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURC...

Страница 92: ...gister can be found under the STATus OPERation command on page 3 33 Unused bits always return a value of 0 Reading the Operation Condition register does not clear it There is no command associated wit...

Страница 93: ...AM otherwise the register value at power on is set to H0000 Executing the STATus PRESet command sets the register value to H0000 STATus OPERation ENABle Mask STATus OPERation ENABle Mask Query respons...

Страница 94: ...by the CLS command There is no command associated with this query Parameter Query response Not applicable NR1 VXI Monitor event word bits Definition D0 IRQ1 occurred D1 IRQ2 occurred D2 IRQ3 occurred...

Страница 95: ...ction Manual Reading the register clears it Query Response STATus OPERation VXIInterrupt CONDition STATus OPERation VXIInterrupt ENABle Dependencies Examples Related Commands Artisan Technology Group...

Страница 96: ...OPERation VXIInterrupt command The enhanced monitor latches the state of the VXI IRQ lines into a hardware register This register is cleared upon VXI Monitor initialization This query returns the stat...

Страница 97: ...etected the query returns a value of 0 in all fields At power on or starting the VXI monitor the values of all fields are set to 0 There is no command associated with this query Parameter Query respon...

Страница 98: ...e at power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus OPERation VXIInterrupt ENABle Mask STATus OPERation VXIInterrupt ENABle Mask Query response NR...

Страница 99: ...ng See VXIMonitor Subsystem on page 3 155 for these events to be recorded There is no command associated with this query Parameter Query response Not applicable NR1 VXI Monitor event word bits Definit...

Страница 100: ...on Manual 3 43 Reading the register clears it Query Response STATus OPERation VXIMonitor CONDition STATus OPERation VXIMonitor ENABle Dependencies Examples Related Commands Artisan Technology Group Qu...

Страница 101: ...ignal assertion time When no ACFAIL is detected the query returns a value of 0 in all fields At power on or restarting VXIMonitor the values of all fields are set to 0 There is no command associated w...

Страница 102: ...ry returns NONE At power on the value is set to NONE There is no command associated with this query Parameter Query response Not Applicable A16 A24 A32 NONE 0 None None Query Response STATus OPERation...

Страница 103: ...ways return a value of zero Definition of the bits in this register can be found in STATus OPERation VXIMonitor command There is no command associated with this query Parameter Query response Not appl...

Страница 104: ...value at power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus OPERation VXIMonitor ENABle Mask STATus OPERation VXIMonitor ENABle Mask Query response NR...

Страница 105: ...ignal assertion time When no SYSFAIL is detected the query returns a value of 0 in all fields At power on or after restarting VXIMonitor the values of all fields are set to 0 There is no command assoc...

Страница 106: ...query Parameter Query response Not applicable NR1 VXI Monitor event word bits Definition D0 TTL0trig occurred D1 TTL1trig occurred D2 TTL2trig occurred D3 TTL3trig occurred D4 TTL4trig occurred D5 TTL...

Страница 107: ...liFrame Mainframe Instruction Manual Query Response STATus OPERation VXITrigger CONDition STATus OPERation VXITrigger ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation...

Страница 108: ...Tus OPERation VXITrigger command The enhanced monitor latches the state of VXI TTL Trigger lines into a hardware register This register cleared upon VXI Monitor initialization This query returns the s...

Страница 109: ...ll fields At power on or after starting the VXI monitor the values of all fields are set to 0 There is no command associated with this query STATus OPERation VXITrigger TRIGger trigger Parameter Query...

Страница 110: ...alue at power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus OPERation VXITrigger ENABle Mask STATus OPERation VXITrigger ENABle Mask Query response NRf...

Страница 111: ...ents The Error Event Queue is preset to the default value of 499 0 The SCPI Positive Transition Filters are preset to 1 This command does not clear the SCPI event registers or the IEEE 488 2 enable an...

Страница 112: ...command There is no command associated with this query Parameter Query response Not applicable NR1 Questionable status word bits Definition D0 VOLTage D1 CURRent D2 Not Used D3 POWer D4 TEMPerature D5...

Страница 113: ...0A IntelliFrame Mainframe Instruction Manual Query Response STATus QUEStionable CONDition STATus QUEStionable ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation Guarante...

Страница 114: ...a value of zero The register is cleared by the CLS command There is no command associated with this query Parameter Query response Not applicable NR1 Blower status word bits Definition D0 Blower 1 D1...

Страница 115: ...truction Manual Reading the register clears it Query Response STATus QUEStionable BLOWer CONDition STATus QUEStionable BLOWer ENABle Dependencies Examples Related Commands Artisan Technology Group Qua...

Страница 116: ...the bits in this register can be found under the STATus QUEStionable BLOWer command on page 3 57 Reading this register does not clear it There is no command associated with this query Parameter Query...

Страница 117: ...E see PSC command on page 3 166 at power on the register value will be restored from nonvolatile RAM otherwise the register value at power on is set to H7FFF Executing the STATus PRESet command sets t...

Страница 118: ...20A There is no command associated with this query Blower Query response BLOWer1 BLOWer2 NR1 0 224 Illegal Parameter Value You attempted to use an invalid parameter value None Query Response STATus QU...

Страница 119: ...er can be found under the STATus QUEStionable command on page 3 55 Reading the register does not clear it There is no command associated with this query Parameter Query response Not Applicable NR1 0 N...

Страница 120: ...d by the CLS command There is no command associated with this query Event Query response Not applicable NR1 Current status word bits Definition D0 24 V D1 12 V D2 5 V D3 Not Used D4 Not Used D5 2 V D6...

Страница 121: ...liFrame Mainframe Instruction Manual Query Response STATus QUEStionable CURRent CONDition STATus QUEStionable CURRent ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation...

Страница 122: ...o command associated with this query Power supply Query response P24 P12 P5 N2 N5PT2 N12 N24 NR2 Not Applicable 224 Illegal Parameter Value You attempted to use an illegal parameter value None Query R...

Страница 123: ...efinition of the bits in this register can be found in STATus QUEStionable CURRent command Reading the register does not clear it There is no command associated with this query Parameter Query respons...

Страница 124: ...wer on is set to H7FFF Executing the STA Tus PRESet command sets the register value to H7FFF STATus QUEStionable CURRent ENABle Mask STATus QUEStionable CURRent ENABle Mask Query response NRf Nondecim...

Страница 125: ...R2 Code Power supply P24 24V P12 12V P5 5V N2 2V N5PT2 5 2V N12 12V N24 24V Not Applicable 224 Illegal Parameter Value You attempted to use an illegal parameter value None Query Response STATus QUESti...

Страница 126: ...es MAX each supply P24 P12 P5 N2 N5PT2 N12 N24 0 0 13 0 13 0 95 0 30 0 60 0 13 0 13 0 Not Applicable 224 Illegal Parameter Value You attempted to set the location to an illegal value 222 Data Out of R...

Страница 127: ...s There is no command associated with this query Power supply Query response Amperes P24 P12 P5 N2 N5PT2 N12 N24 NR2 Not Applicable 224 Illegal Parameter Value You attempted to use an illegal paramete...

Страница 128: ...alue will be restored from nonvolatile RAM otherwise the register value at power on is set to H0000 Executing the STATus PRESet command sets the register value to H0000 Mask Query response NRf Nondeci...

Страница 129: ...00 hours Time Query response NRf 1 N 65535 MIN MAX NR1 1 65535 0 222 Data Out of Range You attempted to set the filter cleaning time interval to an illegal value None Command Query Response STATus QUE...

Страница 130: ...r the next filter cleaning cycle Parameter Query response 0 NR1 0 0 None Query Response STATus QUEStionable FILTer TINTerval STATus QUEStionable FILTer TRESet Syntax Parameters Reset Value Errors and...

Страница 131: ...n the front panel display when the front panel displays the filter cleaning warning message or the filter cleaning remaining time Option 1M only Parameter Query response Not Applicable Not Applicable...

Страница 132: ...here is no command associated with this query Event Query response Not applicable NR1 Current status word bits Definition D0 Not Used D1 Not Used D2 Not Used D3 Not Used D4 Not Used D5 Not Used D6 Not...

Страница 133: ...elliFrame Mainframe Instruction Manual Query Response STATus QUEStionable POWer CONDition STATus QUEStionable POWer ENABle Examples Related Commands Artisan Technology Group Quality Instrumentation Gu...

Страница 134: ...on the returned value is in watts There is no command associated with this query Power supply Query response Not Applicable NR2 Not Applicable None None Query Response STATus QUEStionable POWer LEVel...

Страница 135: ...n of the bits in this register can be found in STATus QUEStion able POWer command Reading the register does not clear it There is no command associated with this query Parameter Query response Not App...

Страница 136: ...power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus QUEStionable POWer ENABle Mask STATus QUEStionable POWer ENABle Mask Query response NRf Nondecimal...

Страница 137: ...iated with this query Power Supply Query response Not Applicable NR2 Not Applicable None None Query Response STATus QUEStionable POWer STATus QUEStionable POWer CONDition STATus QUEStionable POWer ENA...

Страница 138: ...mit Query response NRf MIN MAX NR2 0 0 VX1410A 925 0 VX1420A 1000 0 Not Applicable 222 Data Out of Range You attempted to set the power limit to an illegal value None Query STAT QUES POW LIM Response...

Страница 139: ...nance was performed the returned value is in watts There is no command associated with this query Power supply Query response Not Applicable NR2 Not Applicable None None Query Response STATus QUEStion...

Страница 140: ...se Not applicable NR1 Blower status word bits Definition D0 Slot 0 Delta Temperature D1 Slot 1 Delta Temperature D2 Slot 2 Delta Temperature D3 Slot 3 Delta Temperature D4 Slot 4 Delta Temperature D5...

Страница 141: ...ion Manual Reading the register clears it Query Response STATus QUEStionable TEMPerature CONDition STATus QUEStionable TEMPerature ENABle Dependencies Examples Related Commands Artisan Technology Grou...

Страница 142: ...DELTa0 DELTa12 AMBient ambient input temperature OUT0 OUT12 output temperature of each slot DELTa0 DELTa12 output temperature minus input temperature of each slot NR1 0 224 Illegal Parameter Value You...

Страница 143: ...QUEStionable TEMPerature command on page 3 83 If the fan speed is already set to the maximum speed and the delta temperature is above the limit for 30 seconds a failure will be reported Reading the re...

Страница 144: ...at power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus QUEStionable TEMPerature ENABle Mask STATus QUEStionable TEMPerature ENABle Mask Query response...

Страница 145: ...temperature OUT0 OUT12 output temperature of each slot DELTa0 DELTa12 output temperature minus input temperature of each slot NR1 0 224 Illegal Parameter Value You attempted to use an illegal paramet...

Страница 146: ...input temperature of each slot NR1 Limit NRf If Location is AMBient 1 Limit 75 MIN MAX If Location is DELTan 5 Limit 30 MIN MAX 1 75 5 30 0 224 Illegal Parameter Value You attempted to set the locatio...

Страница 147: ...OUT0 OUT12 DELTa0 DELTa12 AMBient ambient input temperature OUT0 OUT12 output temperature of each slot DELTa0 DELTa12 output temperature minus input temperature of each slot NR1 0 224 Illegal Paramet...

Страница 148: ...ro The register is cleared by the CLS command There is no command associated with this query Parameter Query response Not applicable NR1 Voltage status word bits Definition D0 24 V D1 12 V D2 5 V D3 5...

Страница 149: ...ruction Manual Reading the register clears it Query Response STATus QUEStionable VOLTage CONDition STATus QUEStionable VOLTage ENABle Dependencies Examples Related Commands Artisan Technology Group Qu...

Страница 150: ...inition of the bits in this register can be found in STATus QUEStionable VOLTage command on page 3 91 Reading the register does not clear it There is no command associated with this query Parameter Qu...

Страница 151: ...e at power on is set to H7FFF Executing the STATus PRESet command sets the register value to H7FFF STATus QUEStionable VOLTage ENABle Mask STATus QUEStionable VOLTage ENABle Mask Query response NRf No...

Страница 152: ...wer supply P24 24 V P12 12 V P5 5 V P5STBY 5 V Standby P5EXT 5 V External N2 2 V N5PT2 5 2 V N12 12 V N24 24 V 0 224 Illegal Parameter Value You attempted to use an illegal parameter value None Query...

Страница 153: ...2 and bits 5 8 of the transition filter are always enabled Bit 15 always returns a 0 The factory default setting is 32743 The 5 V Standby and 5 V External are not monitored STATus QUEStionable VOLTag...

Страница 154: ...s is derived from the standard SCPI command The SYSTem ER Ror query is an alias There is no command associated with this query Parameter Query response Not Applicable NR1 string 0 None None Query Resp...

Страница 155: ...en you send the CLS command The SYSTem ERRor ALL query is an alias There is no command associated with this query Parameter Query response Not Applicable NR1 string 0 None None Query Response 221 Sett...

Страница 156: ...or Event queue is cleared at power on or when you send the CLS command There is no command associated with this query Parameter Query response Not Applicable NR1 0 None None Query Response STATus QUEu...

Страница 157: ...an alias The Error Event queue is cleared at power on or when you send the CLS command There is no command associated with this query Parameter Query response Not Applicable NR1 0 None None Query Res...

Страница 158: ...alias The Error Event queue is cleared at power on or when you send the CLS command There is no command associated with this query Parameter Query response Not Applicable NR1 0 None None Query Respons...

Страница 159: ...power on or when you send the STATus PRESet command the contents of the register is set to 499 0 Event list Query response NRf numeric range NRf numeric range NRf numeric range NRf numeric range nume...

Страница 160: ...r Event queue enable to the value of 1999 1000 The SYS Tem ERRor ENABle ALL command is an alias 0 None None Command STATus PRESet STATus QUEue ENABle CLS Syntax Reset Value Errors and Events Dependenc...

Страница 161: ...ing the register does not clear it There is no command associated with this query Summary information for current power triggers and IRQ reside in the overflow register If bit 30 is set the overflow r...

Страница 162: ...ition and IRQ condition Definition of the bits can be found in STATus S2EVent query on page 3 108 There is no command associated with this query Parameter Query response Not Applicable NR1 0 None None...

Страница 163: ...ry information for current power triggers and IRQ reside in an overflow register If bit 30 is set the overflow register should be queried for more information See the STATus S2EVent query on page 3 10...

Страница 164: ...20 5 Standby D21 5 V External D22 2 V D23 5 2 V D24 12 V D25 24 V D26 BERR D27 SYSFAIL D28 ACFAIL D29 AS D30 Check overflow register D31 Not Used 0 None None Query Response STATus SCONdition STATus S2...

Страница 165: ...a value of zero upon power on or when you send the CLS command Reading the events by this command will not clear the events There is no command associated with this query Parameter Query response Not...

Страница 166: ...L5trig occurred D23 TTL6trig occurred D24 TTL7trig occurred D25 Not Used D26 Not Used D27 Not Used D28 Not Used D29 Not Used D30 Not Used D31 Not Used 0 None None Query Response STAT QUES CURR STAT QU...

Страница 167: ...STATus Subsystem 3 110 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 168: ...in the IntelliFrame Mainframe These default settings do not change when you issue the SYSTem FACTory command Table 3 14 Default factory RS 232 communication settings Command Factory default setting S...

Страница 169: ...Ceive BAUD Baud Rate Query response NRf 300 600 1200 2400 4800 9600 19200 MINimum MAXimum NR1 300 600 1200 2400 4800 9600 19200 300 19200 Not Applicable 222 Data Out of Range You attempted to set the...

Страница 170: ...Bits Query response NRf 7 8 MINimum MAXimum NR1 7 8 7 8 Not Applicable 222 Data Out of Range You attempted to set the bits per character value to an illegal value None Command Query Response All SYST...

Страница 171: ...ine high and ignores the incoming CTS line A value of IBFull or RFR specifies that the two lines are used for hardware flow control Flow Control Query response OFF ON IBFull RFR OFF ON IBF Not Applica...

Страница 172: ...ncoming characters are echoed back to the RS 232 port Boolean Query response NRf N 0 0 ON OFF NR1 1 0 1 0 Not Applicable None None Command Query Response All SYSTem COMMunicate SERial commands Syntax...

Страница 173: ...rror Event queue and will not be output to the RS 232 port directly If the mode is ON all asynchronous error messages are placed in the Error Event queue The event registers will not be updated by any...

Страница 174: ...ify the input before the new line character is received For example for a backspace character use H for a line delete character use U If you enable the line buffer mode the output character n 0x0A wil...

Страница 175: ...nd receive values Avoid using XON pacing when transferring binary data Flow Control Query response NONE XON NONE XON Not Applicable 224 Illegal Parameter Value You attempted to set the flow control to...

Страница 176: ...oth the transmit and receive values Type Query response EVEN NONE ODD EVEN NONE ODD Not Applicable 224 Illegal Parameter Value You attempted to set the parity to an illegal value None Command Query Re...

Страница 177: ...ly in the RAW mode There are no associated queries with these commands Table 3 15 lists the preset values for each PRESet command Table 3 15 Effects of PRESet on serial port parameters SERial port par...

Страница 178: ...nd sets both the transmit and receive values Stop Bits Query response NRf 1 2 MINimum MAXimum NR1 1 2 1 2 Not Applicable 222 Data Out of Range You attempted to set terminator to an illegal value None...

Страница 179: ...across power down and will be reset at power up date Query response NRf NRf NRf year month day Not Applicable 222 Data Out of Range You attempted to set the date to an illegal value None Query Respons...

Страница 180: ...ce date There is no command associated with this query Parameter Query response None year month day Not Applicable None None Query Response SYSTem TIME LMAintenance Syntax Parameters Reset Value Error...

Страница 181: ...g the last item from the queue Parameter Query response Not Applicable NR1 string Not Applicable 350 Queue Overflow The Error Event queue overflowed due to the execution errors of other com mands The...

Страница 182: ...rror Event queue is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query response Not Applicable NR1 string Not Applicable None None Query...

Страница 183: ...vent queue is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query response Not Applicable NR1 Not Applicable None None Query Response STAT...

Страница 184: ...ALL command The Error Event queue is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query response Not Applicable NR1 Not Applicable None...

Страница 185: ...The Error Event queue is cleared at power up upon receipt of the CLS command There is no associated command for this query Parameter Query response Not Applicable NR1 Not Applicable None None Query Re...

Страница 186: ...true or when you send the STATus PRESet command the contents of the register is set to 499 0 Event List Query response NRf numeric range NRf numeric range NRf numeric range NRf numeric range numeric...

Страница 187: ...able to the value of 1999 1000 The STATus QUEue ENABle ALL command is an alias Not Applicable None None Command STATus PRESet STATus QUEue ENABle STATus QUEue ENABle ALL Syntax Reset Value Errors and...

Страница 188: ...PERation VXITrigger ENABle 32767 STATus QUEStionable BLOWer ENABle 32767 STATus QUEStionable CURRent ENABle STATus QUEStionable CURRent LIMit 32767 STATus QUEStionable ENABle 0 STATus QUEStionable FIL...

Страница 189: ...Manual Parameter Query response Not Applicable Not Applicable Not Applicable None None Command None Parameters Reset Value Errors and Events Dependencies Examples Related Commands Artisan Technology...

Страница 190: ...with this query Command Header Query response string string Not Applicable The query returns a null string if the Command Header is not a valid command header being recognized by the IntelliFrame Main...

Страница 191: ...d the command from the VXI interface the interface event message is placed in the RS 232 Error Event queue as part of an event message The message in the Error Event queue looks similar to the followi...

Страница 192: ...determined upon power up based on the configuration of the Enhanced Monitor board Parameter Query response Not applicable VX1410A VX1411A VX1420A Not Applicable None None Query Response IDN Syntax Pa...

Страница 193: ...STATus QUEStionable VOLTage ENABle STATus OPERation VXIMonitor ENABle STATus QUEStionable VOLTage PTR STATus OPERation VXITrigger ENABle SYSTem SYSReset STATus QUEStionable BLOWer ENABle VXIMonitor TI...

Страница 194: ...n VXITrigger ENABle SYSTem COMMunicate SERial ERESponse STATus QUEStionable BLOWer ENABle SYSTem COMMunicate SERial LBUFfer STATus QUEStionable CURRent ENABle SYSTem COMMunicate SERial PACE STATus QUE...

Страница 195: ...ystem 3 138 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Command SYSTem NVRecall Examples Related Commands Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www ar...

Страница 196: ...nse NRf N 0 0 ON OFF NR1 1 0 1 0 Not Applicable 224 Illegal Parameter Value You attempted to set the parameter to an illegal value External 5V power to operate the enhanced monitor when the mainframe...

Страница 197: ...rs for all VX1420A mainframes and for the VX1411A or VX1410A mainframes that have been upgraded to the Super Cooler field upgrade kit VXF20A Parameter Query response Not applicable NOOPT S Not Applica...

Страница 198: ...inframe power has been cycled since the mainframe was manufactured Parameter Query response Not Applicable NR1 Not Applicable None None Query Response None Syntax Parameters Reset Value Errors and Eve...

Страница 199: ...nitor is from the mainframe power supply or from an external power supply Parameter Query response Not Applicable MAIN EXTERNAL Not Applicable None None Query Response None Syntax Parameters Reset Val...

Страница 200: ...0 NOTE If SYSRESET is enabled and the mainframe is powered down commu nication through the RS 232 port is not possible even if an external 5 V is supplied to the enhanced monitor card The display will...

Страница 201: ...tor asserting this command will also reset the Enhanced Monitor When disabled asserting this command will only initialize the Enhanced Monitor VXI interface all other Enhanced Monitor functions are no...

Страница 202: ...d across power down At power up the time will be reset to 0 0 0 Parameter Query response NRf NRf NRf hour minute second Not Applicable 222 Data Out of Range You attempted to set the time to an illegal...

Страница 203: ...erating since the last factory maintenance There is no command associated with this query Parameter Query response None hour minute second Not Applicable None None Query Response SYSTem TIME ON Syntax...

Страница 204: ...ng since power on There is no command associated with this query Parameter Query response Not Applicable hour minute second Not Applicable None None Query Response SYSTem TIME LMAintenance Syntax Para...

Страница 205: ...iFrame Mainframe complies There is no command associated with this query Parameter Query response Not Applicable NR2 1995 0 Not Applicable None None Query Response IDN Syntax Parameters Reset Value Er...

Страница 206: ...s 1010 ADC and DAC Control 1 Yes Yes Yes 1100 Front Panel Display Control Test 1 25 No No Yes 1110 Front Panel Display Control Test 2 36 No No Yes 1120 Front Panel Display Control Test 3 6 No No Yes A...

Страница 207: ...display pixels H Front Panel Display Control Test 2 This test writes the following ASCII characters to the display 0x20 to 0x7E These characters are followed by and C H Front Panel Display Control Tes...

Страница 208: ...no failures The command form executes the same tests but returns no results code Each internal self test has a unique numeric identifier in the range of 1000 to 1999 see Table 3 19 on page 3 149 TST...

Страница 209: ...re are no failures Each internal self test has a unique numeric identifier in the range of 1000 to 1999 see Table 3 19 on page 3 149 Parameter Query response NRf Non decimal number 1000 N 1999 NR1 0 1...

Страница 210: ...ere were no failures There is no associated command for this query Each internal self test has a unique numeric identifier in the range of 1000 to 1999 see Table 3 19 on page 3 149 Parameter Query res...

Страница 211: ...d is reported A value of zero is returned if there were no failures There is no associated command for this query Parameter Query response Not Applicable string error number NR1 0 1000 N 1999 verbose...

Страница 212: ...STAT OPER VXII After operating the monitor it is recommended that the summary register be queried to check if any of these events occurred If the bit is set then the corresponding register should be...

Страница 213: ...ed STAT OPER VXIM The system response 3 BERR and SYSFAIL are set 7 STAT OPER VXIM BERR The system response A16 Bus Error occurred in A16 address space 8 STAT OPER VXIM SYSF The system response 1 30 0...

Страница 214: ...Monitor monitors the VXI bus for the specified time interval The factory default setting is OFF 0 Boolean Query response NR1 N 0 0 ON OFF NR1 1 0 1 0 Not Applicable 221 Settings Conflict You attempted...

Страница 215: ...nning Boolean Query response NR1 N 0 0 ON OFF NR1 1 0 1 0 0 213 Init Ignored You sent INITiate while the VXI Monitor was already monitoring the VXI bus None Command Query Response VXIMonitor TINTerval...

Страница 216: ...y default setting is 10 seconds time Query response NRf 1 N 2147483647 MIN MAX NR1 1 2147483647 Not Applicable 222 Data Out of Range You attempted to set the interval to an illegal value 221 Settings...

Страница 217: ...VXIMonitor Subsystem 3 160 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 218: ...rs the SCPI and IEEE 488 2 event registers and the Error Event queue There is no query associated with this command SCPI and IEEE 488 2 enable registers are not cleared by this command None Not Applic...

Страница 219: ...meric 0 N HFF NR1 Standard status byte definition Query response D0 D1 D2 D3 D4 D5 D6 D7 Operation Complete Not Used Query Error Device Dependent Error Execution Error Command Error Not Used Power On...

Страница 220: ...ts Parameter Query response Not Applicable NR1 Standard status byte definition Query response D0 D1 D2 D3 D4 D5 D6 D7 Operation Complete Not Used Query Error Device Dependent Error Execution Error Com...

Страница 221: ...the id message includes both the SCPI and firmware version numbers Parameter Query response Not Applicable Syntax Defined by IEEE 488 2 Manufacturer Model Number Serial Number Firmware Version or Equi...

Страница 222: ...d Event Status register after the previous command has completed execution The query causes an ASCII one to be placed in the output queue after the previous command has completed execution None Not Ap...

Страница 223: ...rd Event Status Enable Register and the SCPI Enable Register settings will be restored to their previous settings Otherwise the settings will be cleared The factory default setting is 0 Value Query re...

Страница 224: ...with this command This command has no effect other than stopping the VXI Monitor The VXI Monitor stops execution while other monitors remain active None Not Applicable None None Command STATus PRESet...

Страница 225: ...n though this is not required by IEEE 488 2 The factory default setting is 0 Mask Query response NRf Non Decimal Numeric 0 N HFF NR1 Standard status byte definition Query response D0 D1 D2 D3 D4 D5 D6...

Страница 226: ...plicable NR1 Standard status byte definition Definition D0 D1 D2 D3 D4 D5 D6 D7 Not Used Not Used Error Event Queue Questionable Event Register MAV ESB Standard Event Status Reg RQS MSS Operation Even...

Страница 227: ...ach internal self test has a unique identifier in the range of 1000 to 1999 TST and TEST are identical queries There is no command associated with this query You can obtain additional test results wit...

Страница 228: ...is no query associated with this command WAI prevents the instrument from executing further commands until the previous command has completed execution None Not Applicable None None Command OPC Synta...

Страница 229: ...IEEE 488 2 Common Commands 3 172 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 230: ...into a 16 bit integer with the most significant bit always zero positive logic Unused or don t care bits in masks and registers always return zeros when queried The Operation Status register and the Q...

Страница 231: ...ble CONDition STATus QUEStionable ENABle Error Event Queue Not Empty Summary of VOLTage FILTer Summary of TEMPerature Summary of BLOWer Summary of VXI Monitor Power On Not Used Command Error Execution...

Страница 232: ...t ENABle Summary of VXI Trigger Monitor IRQ5 occurred IRQ6 occurred IRQ7 occurred Figure 3 12 VXI Interrupt Monitor status register 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 VXI Monitor Status Registers N...

Страница 233: ...r within the mainframe for example one of the power supplies is not working properly you could program the mainframe to warn you of possible problems There are five status registers that provide summa...

Страница 234: ...occurs the corresponding bit will be set in the register 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 CURRent Status Not used STATus OPERation CURRent EVENt STATus OPERation CURRent CONDition STATus OPERatio...

Страница 235: ...g the corresponding bit in the Temperature Status register If a problem occurs the corresponding bit will be set in the register 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 TEMPerature Status Registers Not...

Страница 236: ...isters and two separate Error Event queues for the two interfaces You can send the commands listed below from the interface to set up only those status and event registers of that interface This allow...

Страница 237: ...ge of 200 and 299 These errors are detected by the IntelliFrame Mainframe execution control block Execution errors can occur when a program data element following a header was evaluated as outside of...

Страница 238: ...Error 203 Command protected 213 Init Ignored The measurement initiation was ignored because another measurement was already in progress 220 Parameter error A program data element related error occurr...

Страница 239: ...l Table 3 20 IntelliFrame Mainframe error and event codes Cont Code Message 420 Query UNTERMINATED 430 Query DEADLOCKED 440 Query UNTERMINATED after indefinite response 100 System Message Artisan Tech...

Страница 240: ...enable the bit 0 in the Questionable Enable register If any problems occur with any of the power supplies bit 0 of the register will be set Example 2 Monitor the Delta Temperature This example shows h...

Страница 241: ...mmand Send the STAT SEVent query and note the returned information For example assume that the query returns a decimal value of 1073741947 0X4000007b Decoding the information indicates that both fans...

Страница 242: ...Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 243: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 244: ...Source Characteristic Description Source Voltage 90 VRMS to 250 VRMS continuous range Source Frequency 115 V Operation 45 to 66 Hz or 360 to 440 Hz 230 V Operation 45 to 66 Hz Power Consumption VX141...

Страница 245: ...4 V 13 A 1 ARMS 12 A 12 V 13 A Negligible2 13 A 5V 95 A 3 A 92 A 2 V 30 A Negligible2 30 A 5 2 V 60 A Negligible2 60 A 12 V 13 A Negligible2 13 A 24 V 13 A 1 A RMS 12 A 5 V Stdby3 0 A 0 A 0 A VX1420A...

Страница 246: ...mV 2 V 100 mV 100 mV 5 2 V 260 mV 156 mV 12 V 600 mV 360 mV 24 V 1 2 V 720 mV n Maximum DC Load Ripple Noise Voltage Ripple Noise 24 V 150 mVpp 12 V 50 mVpp 5 V 50 mVpp 2 V 50 mVpp 5 2 V 50 mVpp 12 V...

Страница 247: ...mVpp Protection Over voltage protection Under voltage protection Over temperature protection Short circuit protection Voltage Readout Readout of 24 V 12 V 5 V 2 V 5 2 V 5 V Standby if present and 5V...

Страница 248: ...ower rear of mainframe VX1420A Bottom rear and lower rear of mainframe Mainframe Exhaust Top rear and top sides of mainframe Cooling Clearance Two inches on both sides and rear of mainframe Pressuriza...

Страница 249: ...module l s 2 5 2 1 5 1 0 5 0 00 0 Pressure across module mm H2O Note 1 Proposed VXI 8 cooling methods fan filter removed fans at full speed and sea level Note 2 These curves represent airflow through...

Страница 250: ...d laboratory use Australian Declaration of Conformity C Tick Mark Complies with Australian EMC Framework under the Standard applicable product class 2064 governing radiated and conducted emissions Imm...

Страница 251: ...N CSA C22 2 No 1010 1 Safety requirements for electrical equipment for measurement control and laboratory use Conditions of Approval Installation CAT II Local level mains wall sockets This instrument...

Страница 252: ...m operating temperature derated 1 C 1000 ft 305 m above 5000 ft 1524 m Nonoperating 40 000 ft 12 190 m Dynamics Random Vibration Three axis 30 minutes total 10 minutes per axis without VXI modules ins...

Страница 253: ...rs implemented in a 5 x 7 dot matrix with cursor underline Table 4 8 Mechanical Characteristic Description Overall Dimensions Mainframe Height with feet 13 65 in 35 00 cm Width 16 7 in 42 55 cm Depth...

Страница 254: ...mount Option 3R 3 lbs 1 4 kg Cable Tray Option 1U 7 lbs 3 2 kg Cable Tray Option 2U 9 lbs 4 1 kg Shipping Weight Instrument package 90 lbs 40 9 kg maximum with no modules installed Options package 38...

Страница 255: ...4 3 Front view and side view of VX1420A Super Cooler 15 00 in 38 0 cm 15 75 in 40 0 cm 17 50 in 44 5 cm 18 9 in 48 0 cm 26 4 in 67 06 cm Min 31 4 in 79 76 cm Max Figure 4 4 Front view and side view o...

Страница 256: ...iew and side view of IntelliFrame Mainframe with Option 2U and Rackmount Option Table 4 9 VXI and Plug Play Characteristic Description VXI Fully compatible with VXIbus System Specifications Rev 1 4 VX...

Страница 257: ...Specifications 4 14 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 258: ...d injury do not perform any servicing other than that stated in the operating instructions unless you are qualified to do so Refer to all Safety Summaries before performing any service WARNING Artisan...

Страница 259: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 260: ...Performance Verification Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 261: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 262: ...onix TDS500A series oscilloscope with a P6101B probe DC Voltage Regulation The DC voltage regulation can be verified by following the steps listed under Functional Check beginning on page 1 25 If you...

Страница 263: ...2 Connect a ground lead less than one inch to the 1X oscilloscope probe P6101B 3 Connect the probe to Pin 1 of the 25 pin connector and connect the ground lead to one of the ground pins listed in Tab...

Страница 264: ...Theory of Operation Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 265: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 266: ...ther circuit boards and modules in the mainframe The front panel On Standby DC switch connects to one of two connectors at the top of the backplane When the switch is connected to J22 the switch contr...

Страница 267: ...r RS232 Passive Monitor Fan Speed Switch 0J4 0P3 0J1 0J2 J23 J22 5 5 J25 J27 J24 0P2 0P1 1P1 2P1 3P1 4P1 5P1 6P1 7P1 8P1 9P1 10P1 11P1 12P1 1P2 2P2 3P2 4P2 5P2 6P2 7P2 8P2 9P2 10P2 11P2 12P2 J16 Slot...

Страница 268: ...1 The Fan Speed switch S3 selects either variable fan speed VAR or full speed FULL VX1420A The blower connects to the Monitor board at J8 Fan2 The Fan Speed switch S3 lets you select either variable f...

Страница 269: ...Theory of Operation 6 4 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 270: ...Adjustment Procedures Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 271: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 272: ...he Microsoft Terminal program The following steps describe the calibration procedure using the CALibration command 1 Place the IntelliFrame Mainframe in a constant temperature environment 2 Remove all...

Страница 273: ...Send the following commands to enable the password and to calibrate the IntelliFrame Mainframe 12 Check that the CAL query returns a value of zero to indicate no error has been detected during the cal...

Страница 274: ...v buttons to scroll through the delta temperature of the slots All delta temperatures should read 0 C 1 C 16 Power down the mainframe 17 This completes the calibration procedure The rest of this secti...

Страница 275: ...response A return value of 0 means that the calibration has passed and a return value between 10 and 22 indicates a calibration failure in slots 0 through 12 Parameter Query response Not Applicable N...

Страница 276: ...this command the IntelliFrame Mainframe will generate an error Password Query response string VX1410 Not Applicable Not Applicable 221 Settings Conflict You attempted to send an invalid password None...

Страница 277: ...command otherwise the Intelli Frame Mainframe will generate an error Password Query response string VX1410A Not Applicable Not Applicable 221 Settings Conflict You attempted to send an invalid passwo...

Страница 278: ...led dis abled Enabled commands return a value of 1 disabled commands return a value of 0 Parameter Query response Not Applicable NR1 1 0 Not Applicable None None Query Response SYSTem PASSword CDISabl...

Страница 279: ...number must start with an alphabet ical character followed by an optional alphabetical character and by a maximum of six digits string Not Applicable 221 Settings Conflict You attempted to execute a c...

Страница 280: ...Maintenance Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 282: ...empting any procedures in this chapter Refer to the Operating Basics chapter for information on the location of controls indicators and connectors used with the mainframe CAUTION Many components withi...

Страница 283: ...sentative Cleaning procedures consist of exterior and interior cleaning of the mainframe and cleaning the fan filter Periodic cleaning reduces instrument breakdown and increases reliability Clean the...

Страница 284: ...emoved from the rear of the mainframe The filters slide out of the rear of the mainframe as shown in Figure 8 1 Fan filter Blower filters VX1410A VX1420A Figure 8 1 Removing the fan blower filters Cle...

Страница 285: ...me WARNING To avoid electric shock always power down the mainframe and disconnect the power cord before cleaning or servicing the mainframe The following tools are required to disassemble the IntelliF...

Страница 286: ...rive screws 6 each side using a 3 32 Allen wrench 2 Lift the front bottom corners up 3 After tilting the cover up about 3 inches lift the cover straight off the instrument Remove screws 12 Step 1 Step...

Страница 287: ...an assembly as shown in Figure 8 5 9 VX1410A Remove the cable tie holding the fan cables to the assembly Then remove the fans from the assembly 10 VX1420A Perform steps 1 through 5 11 VX1420A Remove t...

Страница 288: ...aptive screw cable cover 8 32 Captive screw fanhousing 8 32 Captive screw fan housing 8 32 Captive screw fanhousing 8 32 Captive screw Enhance monitor 8 32 Captive screw fan housing Figure 8 3 Locatio...

Страница 289: ...assembly Power supply Fan cables Cable cover Figure 8 4 VX1410A Removing the fan assembly Cable tie Fans Remove screws 6 Remove screws 2 Filter cover Figure 8 5 VX1410A Removing the fans Artisan Techn...

Страница 290: ...r supply Fan cable Cable cover Figure 8 6 VX1420A Removing the blower assembly Cable tie Blower assembly Blower Screws 2 Screws 4 Screws 6 Screws 4 Chassis Shroud Figure 8 7 VX1420A Removing the blowe...

Страница 291: ...o calibrate the main frame Refer to Adjustment Procedures beginning on page 7 1 for information on calibrating the mainframe Refer to Table 8 1 to ensure J2 J38 and J19 jumper configurations are match...

Страница 292: ...e Instruction Manual 8 11 Monitor board Remove fan cables Remove cable cover 8 32 Captive screw 8 32 Captive screws Figure 8 8 Removing the Enhanced Monitor Artisan Technology Group Quality Instrument...

Страница 293: ...rd guides at the top and bottom of the mainframe are very similar The main difference is that the bottom guides IntelliGuides include the spring loaded shutters to redirect air into the mainframe The...

Страница 294: ...uides are replaced as a unit The IntelliGuides are not intended to be disassembled To replace a card guide slide the card guide towards the rear of the mainframe and allow the front of the card guide...

Страница 295: ...inframe allow the front of the card guide to snap into place Gently pull the card guide forward until it pops out of place Installing card guides Removing card guides Front Rear Figure 8 10 Removing t...

Страница 296: ...re 1 on page 8 5 to remove the mainframe cover 1 Loosen all module retaining screws it may be necessary to slide the modules partially out of the mainframe 2 Slide the nut rail out of hole on the side...

Страница 297: ...n place 3 Tilt the board forward and lift the board out of the holes at the top of the mainframe being careful not to damage any components on the circuit board NOTE If you replace the Temperature Sen...

Страница 298: ...d disconnect the ribbon cable from the rear of the Display assembly 3 Remove the 6 32 screws holding the Display assembly in place 4 Gently remove the Display assembly from the mainframe When you reco...

Страница 299: ...gure 8 14 2 Remove two 4 40 Torxdrive T 10 screws that secure each EMI DIN shield to the backplane 3 Remove EMI DIN shield from the backplane Install the EMI DIN shield by reversing the disassembly pr...

Страница 300: ...at J22 or from J23 at the top of the backplane circuit board 4 Disconnect the SysReset cable if attached at J29 at the top of the backplane circuit board 5 From the rear of the mainframe remove the fi...

Страница 301: ...telliFrame Mainframe Instruction Manual 6 32 Screws 17 Connector 0J4 J22 J23 J29 J24 Power switch connector Figure 8 15 Removing the backplane Artisan Technology Group Quality Instrumentation Guarante...

Страница 302: ...ame Table 8 2 summa rizes the different fuses that can be used as the line fuse in the mainframe Table 8 2 Fuses Voltage Rating Tektronix replacement part number 103 V to 250 V operation 15 A Fast Blo...

Страница 303: ...ch fan is properly connected to the monitor board see Figure 8 4 on page 8 8 H Check that both fans are working at their proper speeds H Verify that all of the empty slots within the mainframe are clo...

Страница 304: ...Replaceable Parts Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 305: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 306: ...on number if applicable If you order a part that has been replaced with a different or improved part your local Tektronix field office or representative will contact you concerning any change in part...

Страница 307: ...ix part number Power Supply Module Power supply 119 5553 XX Air Filter VX1410A Air filter Woven Polypropylene black 0 125 thick 2 layer vinyl edging 378 2075 00 Air Filter VX1420A Air filter Woven Pol...

Страница 308: ...ffective Column four indicates the serial number at which the part was discontinued No entries indicates the part is good for all serial numbers 5 Qty This indicates the quantity of parts used 6 Name...

Страница 309: ...OMPANY INDUSTRIAL TAPE DIVISION 3M CENTER ST PAUL MN 55144 1000 01295 TEXAS INSTRUMENTS INC SEMICONDUCTOR GROUP 13500 N CENTRAL EXPRESSWAY PO BOX 655303 DALLAS TX 75272 5303 04713 MOTOROLA INC SEMICON...

Страница 310: ...Replaceable Parts VX1410A VX1420A IntelliFrame Mainframe Instruction Manual 9 5 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 311: ...00 1 CHASSIS VXI MAINFRAME SHEET METAL ASSY INCLUDES ITEMS 3 THROUGH 5 80009 441 2115 00 4 348 1542 00 4 FOOT CABINET BLACK RUBBER 74594 348 1542 00 5 212 0204 00 4 SCR ASSY WSHR 8 32 X 0625 L PNH PL...

Страница 312: ...arts VX1410A VX1420A IntelliFrame Mainframe Instruction Manual 9 7 1 2 3 4 5 4 5 Figure 9 1 Cabinet and chassis assembly Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www a...

Страница 313: ...17 00 1 BACKPLANE VXI BACKPLANE VXI COMPATIBLE WITH 13 C SIZE SLOTS ELECTRONIC AUTOMATIC 80009 118 9417 00 7 020 2194 00 1 COMPONENT KIT BACKPLANE EMI SHIELD KIT CONTAINS DIN SHIELD WITH 2 SCREWS ITEM...

Страница 314: ...1420A IntelliFrame Mainframe Instruction Manual 9 9 1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Figure 9 2 Circuit boards and chassis parts Artisan Technology Group Quality Instrumentation Guaranteed 888...

Страница 315: ...EXT RLV 8 32 X 0 375 BUTTON HEAD HEX DRIVE STAINLESS STEEL BLACK OXIDE FINISH 0 093 DRIV 0KB01 212 0193 00 10 211 0512 00 6 SCREW MACHINE 6 32 X 0 5 FLH 100 DEG STL POZ 73893 ORDER BY DESCRIPTION 11...

Страница 316: ...lliFrame Mainframe Instruction Manual 9 11 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 7 18 Figure 9 3 Power supply monitor and VX1410A fan assembly Artisan Technology Group Quality Instrumentation Guar...

Страница 317: ...1 COVER FAN CABLE COVER 0 040 A1 ALLOY TK1943 OBD 200 4344 00 3 119 5199 00 1 FAN DC BLOWER 48V DUAL INLET 450 CFM 177 W WITH 6 SCREWS 80009 119 5199 00 4 212 0400 00 4 SCREW MACHINE 8 32 X 0 250 PNH...

Страница 318: ...10A VX1420A IntelliFrame Mainframe Instruction Manual 9 13 1 2 3 4 5 6 1 3 6 5 Figure 9 4 VX1420A or Option VXF20A blower assembly Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SO...

Страница 319: ...INE 8 32 X 0 5 FLH 100 DEG STL CDPL T 15 0KB01 ORDER BY DESC 3 407 4524 00 1 BRACKET LEFT RACKMOUNT SILVER GRAY TK1943 407 4524 00 4 212 0671 00 4 SCREW MACHINE 10 32 X 0 625 FLH 100 DEG STL CD PL TOR...

Страница 320: ...s VX1410A VX1420A IntelliFrame Mainframe Instruction Manual 9 15 1 2 3 4 2 2 2 7 6 5 6 7 8 Figure 9 5 1R Rackmount assembly Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE ww...

Страница 321: ...T RACKMOUNT SILVER GRAY TK1943 407 4524 00 5 211 0718 00 8 SCREW MACHINE 6 32 X 0 312 FLH100 STL CDPL T 10 0KB01 211 0718 00 6 212 0671 00 4 SCREW MACHINE 10 32 X 0 625 FLH 100 DEG STL CD PL TORX DRIV...

Страница 322: ...A IntelliFrame Mainframe Instruction Manual 9 17 1 2 3 4 5 6 7 6 10 1 5 2 2 2 11 3 8 8 9 10 11 12 Figure 9 6 2R Rackmount and door assembly Artisan Technology Group Quality Instrumentation Guaranteed...

Страница 323: ...X 1 COMPONENT KIT CABLE TRAY 1U KIT KIT INCLUDES ITEMS 1 THRU 3 80009 0202223XX 1 212 0158 00 14 SCREW MACHINE 8 32 X 0 375 PNH STL CDPL T 20 0KB01 ORDER BY DESC 2 333 4158 00 1 PANEL CONN 0 09ALUM 80...

Страница 324: ...eable Parts VX1410A VX1420A IntelliFrame Mainframe Instruction Manual 9 19 1 2 3 4 Figure 9 7 Option 1U cable tray Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisan...

Страница 325: ...2U KIT KIT INCLUDES THE ITEMS 1 THRU 4 80009 0202224XX 1 426 2537 00 2 RAIL FOOT CHASSIS ALUM SHEET METAL TK1943 426 2537 00 2 212 0158 00 16 SCREW MACHINE 8 32 X 0 375 PNH STL CDPL T 20 0KB01 ORDER B...

Страница 326: ...able Parts VX1410A VX1420A IntelliFrame Mainframe Instruction Manual 9 21 1 2 3 4 5 Figure 9 8 Option 2U cable tray Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisa...

Страница 327: ...PT 1M 80009 333 4241 00 2 119 5086 00 1 DISPLAY MODULE VACUUM FLOURESCENT 1 X 20 ALPHANUMERIC 5 X 7 DOT MATRIX W CURSOR 4T165 FC20X1RA AA AB 3 671 3290 00 1 CKT BD ASSY FRONT PANEL DISPLAY 80009 671 3...

Страница 328: ...s VX1410A VX1420A IntelliFrame Mainframe Instruction Manual 9 23 1 2 3 4 5 6 7 Figure 9 9 Display panel VX1420A or Option 1M Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE w...

Страница 329: ...Replaceable Parts 9 24 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 330: ...Appendices Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 331: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 332: ...ers of the IntelliFrame Mainframe and any installed modules Enter the steady state current IMP and the dynamic current IMD for each module Add the individual currents to determine the total current ne...

Страница 333: ...Power Supply Current Limit Current Limits Power Calculation Total Power 925W 1410A 95A 14 5A 13A 3A 13A 3A 13A 9A 13A 9A 60A 12A 30A 7A Individual Current Sums Less Than Power Supply Total Current Ext...

Страница 334: ...the configuration procedures for the mainframe described in this appendix Remote Power Switch Configuration You can control the power on function of the mainframe by the front panel On Standby switch...

Страница 335: ...cable at J22 and connect it to J23 5 Replace the mainframe cover 6 Connect the momentary switch to pins 5 and 18 return side of the 25 pin rear panel connector J22 J23 J29 J24 Power switch connector...

Страница 336: ...set feature can be utilized Perform the following procedures to activate the SYSReset feature 1 Remove the mainframe cover 2 Cut the tie wrap that secures the SYSReset connector to the front panel swi...

Страница 337: ...dby voltage from the 5 V external supply Monitor Backup Power Supply connector then the external supply will provide the backplane with the 5 V external voltage 1 A maximum If you source the 5 V Stand...

Страница 338: ...andard or not Table C 1 SCPI conformance information Command Defined in SCPI 1995 0 Not de fined In SCPI 1995 0 CALibration ALL n DISPlay WINDow TEXT DATA n MARQuee n STATus OPERation EVENt n CONDitio...

Страница 339: ...me Instruction Manual Table C 1 SCPI conformance information Cont Command Not de fined In SCPI 1995 0 Defined in SCPI 1995 0 n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n Artisan Tec...

Страница 340: ...nstruction Manual C 3 Table C 1 SCPI conformance information Cont Command Not de fined In SCPI 1995 0 Defined in SCPI 1995 0 n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n n Artisan Tec...

Страница 341: ...n FACTory n HELP SYNTax n IEMessage n MODel n NVRecall n NVSave n OPTions n PASSword CDISable n CENable n STATe n POWer n CYCLe n SOURce n SYSReset n MONitor n IMMeditate n TIME LMAintenance n ON n V...

Страница 342: ...ntelliFrame Mainframe Instruction Manual C 5 Table C 1 SCPI conformance information Cont Command Not de fined In SCPI 1995 0 Defined in SCPI 1995 0 n n n n n n n n n Artisan Technology Group Quality I...

Страница 343: ...Appendix C SCPI Conformance Information C 6 VX1410A VX1420A IntelliFrame Mainframe Instruction Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 344: ...Glossary Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 345: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 346: ...same size as a VMEbus single height short module Asynchronous Communication Communications that occur outside the normal command response cycle Such communications have higher priority than synchronou...

Страница 347: ...mum system timing skew of 2 ns and a maximum total delay of 8 ns Commander In the VXIbus interface a device that controls another device a servant A commander may be a servant of another commander Com...

Страница 348: ...evice that is not defined in the VXIbus specification D Size Card A VXIbus instrument module that is 340 0 366 7 mm 30 48 mm 13 4 14 4 in 1 2 in DTB See Data Transfer Bus DTB Arbiter A functional modu...

Страница 349: ...me Mainframe A family of VXI mainframes from Tektronix that automatically direct airflow to the installed modules have an autoconfigure backplane have adjustable rack kits and cable trays optional and...

Страница 350: ...ess time Message A series of data bytes that are treated as a single communication with a well defined terminator and message body Message Based Device A VXIbus device that supports VXI configuration...

Страница 351: ...vides configuration management services such as address map configuration determining system hierarchy allocating shared system resources performing system self test diagnostics and initializing syste...

Страница 352: ...rough 12 in a mainframe The delay between slots is less than 5 ns and the lines are well matched for timing skew STST STart STop protocol used to synchronize modules Super CoolerT This is a IntelliFra...

Страница 353: ...and communication VXIbus Subsystem One mainframe with modules installed The installed modules include one module that performs slot 0 functions and a given complement of instrument modules The subsyst...

Страница 354: ...Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 355: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 356: ...1 RS 232 3 4 cooling 1 2 4 5 fault isolation 8 21 8 22 cooling curve 4 6 D device specific error 3 180 diagnostics 8 21 disassembly procedures 8 4 8 20 tools required 8 4 display assembly removing 8 1...

Страница 357: ...guidelines 1 20 monitor backup power supply 2 4 monitor backup power supply connector B 4 monitor board enhanced 6 3 removing 8 10 8 11 monitor display 2 2 monitor display assembly removing 8 17 N nu...

Страница 358: ...Monitor ACFail TIME 3 40 3 52 STATus OPERAtion VXIMonitor BERR ADDRess 3 45 STATus OPERAtion VXIMonitor ENABle 3 41 3 47 3 53 STATus OPERAtion VXIMonitor SYSFail TIME 3 48 STATus PRESet 3 54 STATus QU...

Страница 359: ...0 SYSTem FACTory 3 131 3 132 SYSTem HELP SYNTax 3 133 SYSTem HRESet 3 144 SYSTem IEMessage 3 134 3 135 SYSTem NVRecall 3 136 SYSTem NVSave 3 137 3 138 SYSTem PASSword CDISable 7 5 SYSTem PASSword CENa...

Страница 360: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 361: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Страница 362: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: