Specification
and Performance Check—
FG 501
Table
2-4 (cont)
Performance
Description
Requirement
Application
Example
Distortion
Analyzer
Frequency
range
from 1 Hz
to at least
600 kHz. Dis
tortion
resolution <0.5%.
Check
sine
wave dis
tortion.
Hewlett-Packard
334A
Dis
tortion
Analyzer or equiv
alent.
50 Ω Feedthrough
Termination
(2)
bnc
connectors.
Steps
1,
2, 3, 5, 6, 8,
and 9.
Tektronix
Part No.
011-0049-01.
600 Ω Feedthrough
Termination
bnc
connectors.
TRIG
OUTPUT Amplitude
check
Tektronix
Part No.
011-0092-00.
50 Ω Coaxial
Cables
(2
ea)
bnc connectors.
All.
Tektronix
Part
No.
012-0057-01.
Adapter
bsm-to-bne.
TRIG
OUTPUT Amplitude
check.
Tektronix
Part No.
103-0036-00.
Adapter
Dual banana
plug-to-bnc
female.
VCF
INPUT check.
Tektronix
Part No.
103-0090-00.
Tee
Connector
bnc
connectors.
Basic
timing
check.
Tektronix
Part
No.
103-0030-00.
10X Attenuator
bnc
connectors
50 Ω impedance.
Square wave checks.
Tektronix
Part
No.
011-0059-02.
PRELIMINARY
PROCEDURE
1. Ensure that the
correct nominal
line selector block
has
been installed on the line selector pins on the power
module
interface
board, and
the
regulating
range selected
includes
the applied line voltage.
Refer to the installation
section
of the power module manual.
2.
Ensure that all test equipment is
suitably adapted
to
the
applied line
voltage.
PERFORMANCE
CHECK
PROCEDURE
1. Dial Alignment
a. Set the FG
501
controls as follows:
FUNCTION
Triangle
AMPL1
Fully cw
OFFSET
2
Midrange and
In
PHASE
In
MULTIPLIER
10
3
FREQ
VERNIER
Fully cw
FREQUENCY
Hz
Near
10
3.
Install
the FG 501 into
the power module and, if
applicable,
install the TM 500-Series test equipment into
the test equipment
power module.
4.
Connect
the equipment
under test and the test
equipment
to a suitable
line voltage source. Turn on all
equipment and allow at least
20 minutes for
the equipment
to
stabilize.
'Below
SN
BJ30000
AMPL
control
Is
labeled
OUTPUT.
Tor
SN
below
B130000
set
OFFSET
to
zero.
b.
Adjust the
oscilloscope vertical for dc coupling at
2 V/div
sensitivity. Set
the time base sweep speed to
.1
ms/div. Set
the triggering controls
to internal
slope
operation.
c.
Connect
the
OUTPUT of the FG 501 through a
50 Ω
coaxial
cable
and
a 50 Ω termination,
with the 50 Ω
termination
at the vertical input of the oscilloscope.
2-6
REV A,
MAR 1979
Содержание FG501
Страница 1: ...Tektronix COMMITTED TO EXCELLENCE FUNCTION GENERATOR FG 501 INSTRUCTION MANUAL...
Страница 5: ...Fig 1 1 FG 501 Function Generator FG 501 REV B JUL 1978...
Страница 43: ...Section 6 FG 501 OPTIONS There are no options for the FG 501 at this time 6 1...
Страница 57: ...IMPONENT LOCATION GRID Below SN B020000 See Parts List for serial number ranges t Located on back of board...
Страница 58: ...PARTS LOCATION G R ID BELOW SN B020000 COMPONENT Below SN BOS...
Страница 59: ...FG I...
Страница 60: ...FG 501 BLOCK DIAGRAM BLOCK D IA G R A M...
Страница 62: ...OCATIONGRID See Parts List for serial number ranges f Located on back of board Alternate location...
Страница 63: ...COMPONENT LOC SN B020000 UP PARTS LOCATION G RID SN B 020000 U P REV G JUN 1978...
Страница 65: ......
Страница 66: ...GENERATOR...
Страница 67: ...FQ 501 R 6 V H JUL 1 118 4 BI Ife Output Am...
Страница 68: ...431 I4 Output Amplifier and Switch Details DEH OUTPUT AMPL SWITCH D ETA ILS...
Страница 69: ...8L innr 9 A ra ios J...
Страница 70: ...ULI91a Power Supply O DEM 50 3 POWER SUPPLY...
Страница 76: ......
Страница 77: ...79 78 FG 501 FUNCTION GENERATOR FIG 1 EXPLODED...