2 Description of User Controls
2.1 Acquisition
2.1.1 Run
Starts the acquisition process
– data is captured (if trigger event occurs) and displayed on the
screen. If “continuous” is selected the acquisition will repeat until “Stop” is pressed. In case of
“single shot” acquisition the scope acquires and displays only one record and then stops if “No
Avg” (no averaging) is selected; otherwise it will acquire the given number of records to average
over and then stop. This means that if you want to do a true single shot capture you need to set
the averaging to “No Avg”.
In datalogger (roll) mode single shot and averaging is not available; in this case the Run button
simply starts the continuous acquisition.
2.1.2 Stop
Stops the acquisition process.
2.1.3 Clear
Clears the waveform.