Adjustment Procedure
Before Adjustments
NOTE.
The voltage references inside the oscilloscope are very stable over time
and should not require routine adjustment.
Before performing the adjustment procedure, do the following:
Read the
Service Safety Summary
and the
General Safety Summary
at the
front of this manual.
Perform the Performance Veri
fi
cation procedures. If the oscilloscope is
operating within speci
fi
cations, you do not need to make any adjustments.
(See the
MSO4000B Series and DPO4000B Series Oscilloscopes
Speci
fi
cations and Performance Veri
fi
cation Technical Reference manual
,
which was provided with your product. This manual is also available at
www.tektronix.com/manuals.)
Before performing the adjustment procedure, you must warm up the
oscilloscope and the oscilloscope calibrator for at least 30 minutes in an
ambient temperature between 20 °C and 30 °C. Adjustments performed before
warm-up or outside this temperature range may result in poor performance or
failure of the adjustment procedure.
Learn how to operate the oscilloscope calibrator. Refer to the user manual,
which can be found at http://us.
fl
uke.com.
Navigating During Adjustments
Continue.
Once you have set the calibrator output signal that is speci
fi
ed
on the oscilloscope screen, push the
OK Do Next Step
side-bezel button to
proceed. Information about setting up the calibrator output signal is provided
after the Factory Adjustment Procedure. (See page 13,
Go back.
Push the
Return to Previous Screen
side-bezel button to go back
to the previous step. Push the button multiple times if you want to go back
multiple steps.
Abort.
Push the
Menu Off
front-panel button on the oscilloscope to cancel the
procedure at any time. This causes the oscilloscope to revert to the previous
calibration constants. The
last factory calibration date
will not be updated.
8
MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual
Содержание DPO4034B
Страница 2: ......
Страница 10: ...General Safety Summary vi MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual...
Страница 12: ...Preface viii MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual...
Страница 30: ...Adjustment Procedure 18 MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual...
Страница 42: ...Maintenance 30 MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual...