Setup/hold triggering uses the setup/hold violation zone to detect when data is unstable too near the time it is clocked. Each time
trigger holdoff ends, the instrument monitors the data and clock sources. When a clock edge occurs, the instrument checks the
data stream it is processing (from the data source) for transitions occurring within the setup/hold violation zone. If any occur, the
instrument triggers with the trigger point located at the clock edge.
Positive settings for both setup and hold times (the most common application) locate the setup/hold violation zone so that it
spans the clocking edge as shown above in the top waveform. The instrument detects and triggers on data that does not become
stable long enough before the clock (setup time violation) or that does not stay stable long enough after the clock (hold time
violation).
Negative settings for setup or hold times skew the setup/hold violation zone to locate it before or after the clocking edge as
shown above in the center and bottom waveforms. The instrument can then detect and trigger on violations of a time range that
occurs before or one that occurs after the clock.
NOTE. Keep the hold-time setting to no more than 2.5 ns less than one-half the clock period (hold time < (period/2) –2.5 ns) or
the instrument cannot trigger (this assumes a 50% duty cycle clock).
State trigger
A state trigger occurs when the inputs to the logic function cause the function to become True or False at the time the clock input
changes state. When you use a state trigger, you define:
■
The precondition for logic input channels 1 through 3
■
The direction of the state change for the clock input, channel 4
■
The Boolean logic function: AND, NAND, OR, or NOR
■
The condition for triggering: the Boolean function becomes True (logic high) or False (logic low)
State trigger logic conditions are summarized in the following table.
Oscilloscope reference
672
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Содержание DPO2304SX
Страница 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Страница 2: ......
Страница 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Страница 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 83: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 51...
Страница 89: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 57...
Страница 100: ...About Tektronix oscilloscopes 68 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 234: ...Horizontal and acquisition setups 202 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 276: ...Mask testing setups 244 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 558: ...Copying and printing 526 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Страница 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 687: ...Figure A Figure B Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 655...
Страница 898: ...Oscilloscope reference 866 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...