Adjustment Procedures
3-12
AWG7000 Series Service Manual
Inter-Channel Skew
Calibration
(AWG7102 and AWG7052)
This calibration is required to ensure that the Ch 1 and Ch 2 output skews are set
within the specified range. This calibration is applied to both analog and marker
outputs.
1.
Install the test hookup and preset the instrument controls:
a.
Hook up the sampling oscilloscope:
Attach the SMA terminations to the Channel 1 Analog (–) Output,
Channel 2 Analog (–) Output, and Marker 1 (–) Output on the front
panel of the arbitrary waveform generator.
Connect an SMA cable from the Channel 1 Analog (+) Output on the
front panel of the arbitrary waveform generator to the CH 1 Input of
the sampling oscilloscope using the SMA attenuator, as shown in
Figure 3-6.
Connect an SMA cable from the Channel 1 Marker (+) Output on the
front panel of the arbitrary waveform generator to the TRIGGER
DIRECT INPUT of the sampling oscilloscope using the SMA
attenuator, as shown in Figure 3-6.
Figure 3-6: Inter-channel skew calibration initial hookup
Equipment required
One sampling oscilloscope
Two 50
Ω
SMA 12 dB attenuators
Two 50
Ω
SMA cables
Three 50
Ω
SMA terminations
SMA termination
Instrument under test
Sampling oscilloscope
SMA attenuator
SMA attenuator
Содержание AWG7000 Series
Страница 4: ......
Страница 10: ...List of Tables vi AWG7000 Series Service Manual...
Страница 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Страница 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Страница 18: ...Preface xiv AWG7000 Series Service Manual...
Страница 19: ...Operating Information...
Страница 20: ......
Страница 23: ...Theory of Operation...
Страница 24: ......
Страница 31: ...Adjustment Procedures...
Страница 32: ......
Страница 51: ...Maintenance...
Страница 52: ......
Страница 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Страница 109: ...Replaceable Parts...
Страница 110: ......
Страница 125: ...Replaceable Parts AWG7000 Series Service Manual 5 15 Figure 5 7 Exploded diagram Power supply assembly 1 2 3 4 5 6 6...
Страница 137: ...Replaceable Parts AWG7000 Series Service Manual 5 27 Figure 5 14 Exploded diagram SMA bracket 1 2 3 5 4 6 8 7...
Страница 139: ...Replaceable Parts AWG7000 Series Service Manual 5 29 Figure 5 15 Exploded diagram Relay unit 11 2 7 4 5 6 3 8 10 9 1 4...