
Getting started
DC accuracy. It is recommended that the self calibration should be performed
along with a periodic check.
NOTE.
If you need to verify that the instrument meets the warranted speci
fi
cations,
do the complete set of performance veri
fi
cation procedures provided in the
Speci
fi
cations and Performance Veri
fi
cation manual.
CAUTION.
Do not power off the instrument while executing self calibration. If the
power is turned off during self calibration, data stored in the internal memory
may be lost.
1.
Push the front-panel
Utility button.
2.
Push the -more- bezel
button.
3.
Push the
Diagnostics/Calibration
bezel button.
4.
Do one of the following:
NOTE.
Before executing
self calibration, ensure that
the ambient temperature is
b20 °C and +30 °C
(+68 °F to +86 °F), and allow
a 20 minute warm-up period.
Execute Diagnostics
:
Push this bezel button to
execute the instrument
diagnostics.
Execute Calibrations
:
Push this bezel button to
execute self calibration.
5.
If the diagnostics or
calibration completes
without any errors, the
message “PASSED” is
displayed.
8
AFG2021 Quick Start User Manual
Содержание AFG2021
Страница 2: ......
Страница 3: ...AFG2021 Arbitrary Function Generator ZZZ Quick Start User Manual xx ...
Страница 6: ......
Страница 98: ...Operating Basics 80 AFG2021 Quick Start User Manual ...
Страница 102: ...Application examples 84 AFG2021 Quick Start User Manual ...