AFG1000 Series Arbitrary Function Generator Programmer Manual
39
Command Descriptions
[SOURce[1|2]]:FREQuency:STARt
This command sets or queries the start frequency of sweep for the specified
channel. This command is always used with the
[SOURce[1|2]]:FREQuency:STOP command. The setting range of start
frequency depends on the waveform selected for sweep. For more information,
refer to the
AFG1000 Series Specifications and Performance Verification
Technical Reference
.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:FREQuency:STARt
{<frequency>|MINimum|MAXimum}
[SOURce[1|2]]:FREQuency:STARt
? {MINimum|MAXimum}
Related Commands
Arguments
<frequency>::=<NRf>[<units>]
where:
<NRf> is the start frequency.
<units>::=[Hz | kHz | MHz]
Returns
<frequency>
Examples
SOURce1:FREQuency:STARt 10kHz
sets the sweep start frequency of CH 1 to 10 kHz.
[SOURce[1|2]]:FREQuency:STOP
This command sets or queries the stop frequency of sweep for the specified
channel. This command is always used with the
[SOURce[1|2]]:FREQuency:STARt command. The setting range of stop
frequency depends on the waveform selected for sweep. For more information,
refer to the
AFG1000 Series Specifications and Performance Verification
Содержание AFG1000 Series