Operating Instructions
—
7B53A/7B53AN
Operators
Display
Modes
Four
Display Modes can be selected by
appropriate
settings
of the Time/Division switches (see
Fig.
1-7).
Main Sweep Operation. To select
the
MAIN SWP Display
Mode
(non-delayed
sweep) set the TIME/DIV
OR DLY
TIME and
DLY'D
Time/Division switches to the same
sweep
rate and press
in the DLY'D
Time/Division switch.
Calibrated sweep
rates in the MAIN SWP Display Mode are
available
from
5 s/DIV
to 0.05 µs/DIV (5 ns/DIV with
X10
magnification).
The VARIABLE
control (Variable Selector
connector
set
to
Main Variable) can
be used to obtain
uncalibrated
sweep
rates to 12.5 seconds/division. Trig
gering
in the MAIN SWP
Display Mode is controlled by the
MAIN TRIGGER
controls.
Intensified
Sweep
Display. To select
the INTEN
Display
Mode, pull
out
the DLY'D Time/Division knob and rotate
it
to a desired sweep
rate faster than the
TIME/DIV OR
DLY
TIME
switch setting. The INTEN Display Mode
provides an intensified
portion of the main sweep during
the
time
the delayed
sweep is in operation (see Fig. 1-10A).
The
time
that the delayed sweep
runs is determined by the
DLY'D Time/Division
switch. The amount of delay
time
between the
start of the delaying sweep and the intensified
portion
is determined by
the TIME/DIV OR DLY TIME
switch
and the DELAY
TIME MULT dial. Triggering for
the
delaying
sweep portion of display is controlled by the
MAIN
TRIGGERING controls; triggering for the intensified
portion
of display is controlled by
the DLY'D TRIG
controls.
Delayed
Sweep Display. The
DLY’D SWP Display Mode
is
selected when
the DLY'D Time/Division switch
is pulled
out,
rotated
in the INTEN Mode for the desired sweep rate,
and then
pushed in. In this mode, the intensified portion of
display, as
viewed
in the INTEN
Display Mode, is displayed
on
the CRT at the sweep rate
indicated by
the DLY'D
Time/Division switch (see
Fig. 1-10B). Calibrated sweep
rates
in the DLY'D SWP Display Mode are available from
.5
s/DIV
to 0.05 µs/DIV. Uncalibrated delayed sweep rates to
1.25
seconds/division
can
be obtained by using
the VARI
ABLE control
(Variable
Selector connector
set to Dly'd
Variable
position). Triggering for the DLY'D SWP display is
controlled
by the DLY'D TRIG controls.
Mixed
Sweep Operation. The MIXED Display Mode is
selected
when
the DLY'D SWP Display Mode
is selected
and
the VARIABLE knob is pulled out. In
this mode, the
main
sweep
is displayed on
the CRT to a point determined
by
the DELAY TIME MULT
dial; the
remainder of the
sweep
is at
a rate determined by the delayed sweep.
Triggering for
the
main sweep portion of display is
controlled by the
MAIN TRIGGERING controls; and
triggering for
the delayed sweep portion
of display is
controlled
by the
DLY'D TRIG controls. A typical mixed
sweep
display is
shown in
Fig.
1-11.
Delay
Time
Multiplier
The DELAY
TIME MULT dial
(functional in the
INTEN, DLY'D
SWP, and MIXED Display Modes) provides
0
to
10 times
continuous sweep delay. The time
that the
main sweep
runs before the
start of the delayed sweep is
determined
by
the settings of the TIME/DIV OR DLY
TIME
switch and
the DELAY TIME
MULT dial.
For example, the
delay time indicated by the DELAY
TIME
MULT dial
in Fig. 1-12 is 3.55; this corresponds to
3.55
CRT divisions of main sweep. Thus 3.55 multiplied by
the
delaying
sweep rate, indicated by the TIME/DIV OR
DLY
TIME switch, gives the calibrated delay time before
the start of the
delayed sweep.
Fig. 1-11. Typical Mixed
Sweep display (DELAY TIME MULT dial
set
to
3.55).
1-13