FlexRay serial bus trigger settings panel
NOTE. Requires option SRAUTO.
Field or control
Description
Source
Select the FlexRay bus on which to trigger.
Trigger On
Select the type of information on which to trigger.
Indicator Bits
Select the indicator bits type on which to trigger.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Indicator Bits or Header Fields.
Identifier
Sets the frame identifier pattern on which to trigger.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Frame ID, Header Fields, or Identifier & Data.
Cycle Count
Sets the cycle count pattern on which to trigger. Use in conjunction with the Trigger When field
to specify the exact trigger condition.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Cycle Count or Header Fields.
Payload Length
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Header Fields.
Header CRC
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Header Fields.
Data Bytes
Sets the number of data bytes on which to trigger (one to sixteen bytes). Use the A knob to
change the value. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data or Identifier & Data.
Data
Sets the data pattern on which to trigger. The number of bits shown depends on the Data Bytes
setting. Use in conjunction with the Trigger When field to specify the exact trigger condition.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data or Identifier & Data.
Byte Offset
Sets the byte offset (Don't Care or the number of bytes). Tap the input box and use the A knob
to change the value.
Available when Trigger On = Data or Identifier & Data.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
353
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...