![Tektronix 59021-10 Скачать руководство пользователя страница 237](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414237.webp)
Appendix B: Performance Verification
BĆ44
DG2040 Data Generator
DG2040 Test Record
Serial Number:
Certificate Number:
Calibration Date:
Technician:
Table BĆ9: DG2040 test record
Clock frequency tests
Low limit
Test result
High limit
Internal clock frequency
acc ac
1.100 000 GHz
1.099 998 9 GHz
1.100 001 1 GHz
accuracy
650.100 0 MHz
650.099 35 MHz
650.100 65 MHz
200.000 0 kHz
199.999 80 kHz
200.000 20 kHz
100.000 0 mHz
99.999 90 mHz
100.000 10 mHz
High level output voltage accuracy (DC Data
Out)
Low limit
Test result
High limit
-0.875 V
-0.951 V
-0.799 V
0 V
-0.050 V
0.050 V
CH0 Output
+1.00 V
0.920 V
1.080 V
+2.00 V
1.890 V
2.110 V
+3.50 V
3.345 V
3.655 V
-0.875 V
-0.951 V
-0.799 V
0 V
-0.050 V
0.050 V
CH0 Output
+1.00 V
0.920 V
1.080 V
+2.00 V
1.890 V
2.110 V
+3.50 V
3.345 V
3.655 V
-0.875 V
-0.951 V
-0.799 V
0 V
-0.050 V
0.050 V
CH1 Output
+1.00 V
0.920 V
1.080 V
+2.00 V
1.890 V
2.110 V
+3.50 V
3.345 V
3.655 V
-0.875 V
-0.951 V
-0.799 V
0 V
-0.050 V
0.050 V
CH1 Output
+1.00 V
0.920 V
1.080 V
+2.00 V
1.890 V
2.110 V
+3.50 V
3.345 V
3.655 V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com