Section 3—5110
ADJUSTMENT
Introduction
This adjustment procedure is to be
used to restore the
oscilloscope
to original
performance specifications. Ad
justment
need
not be performed unless the instrument
fails
to meet the requirements listed in the
Specification
section
of this manual, or the
Performance Check cannot
be
completed
satisfactorily.
Completion
of all adjustment steps in this procedure
ensures
that
the
instrument will meet the performance
requirements
listed in the Specification section. However,
to fully
ensure
satisfactory performance, it is recommend
ed
that
the Performance Check be performed after any
adjustment
is
made.
Tektronix
Field Service
Tektronix,
Inc. provides
complete instrument repair
and
recalibration at
local Field Service
Centers and the
Factor
Service Center. Contact your local
Tektronix
Field
Office
or
representative
for further information.
Test
Equipment Required
The following
test
equipment, or equivalent, is required
for
complete adjustment of the oscilloscope. The test
equipment
performance
requirements
listed are the
minimum necessary
for accurate adjustment. Substitute
equipment
must meet or exceed the stated requirements.
All
test equipment is
assumed to be operating within
tolerance.
Table
3-1
LIST
OF TEST EQUIPMENT REQUIREMENTS FOR
ADJUSTMENT
Description
Performance
Requirements
Application
Example
Amplifier
plug-in
Bandwidth, de
to 2 MHz;
deflection
factor, 5 mV
to
5 V/dlv.
One
required for all
tests. Two
re
quired
for step 13.
a. TEKTRONIX 5A15N
Amplifier.
b.
TEKTRONIX
5A18N
Dual-Trace
Amplifier.
Time-base plug-in
unit
Sweep rate, at least
2
µs/div.
Steps
1 through 12.
a. TEKTRONIX 5B1
ON
Time-Base.
Calibration generator
Amplitude
calibration,
5 mv
to 5 V;
accuracy,
±0.25% into 1
MΩ; output,
square wave
at approximately
1
kHz.
Steps
10 and 12.
a.
TEKTRONIX
PG 506
Calibration
Generator.
Sine-wave generator
Frequency, 100
kHz;
output
amplitude,
variable
from
250 mV to 6 V into 50 Ω.
Step
13.
a. TEKTRONIX
FG 503
Function
Generator.
Digital
voltmeter
Range, zero
to
250 volts;
accuracy,
within
0.1%.
Steps 1 through
4.
a.
TEKTRONIX DM
501
Digital
Multimeter.2
DC
voltmeter (vom)
Range,
zero to 4000 volts;
accuracy, checked to
within
1%
at 3400 volts.
Step
5.
a.
Triplett
Model
630-NA.
b.
Simpson
Model 262.
Coaxial cable
Impedance, 50 Ω;
length,
42
inch;
connectors,
bnc.
Steps
10,
12, and
13.
a. TEKTRONIX
part
012-0057-01.
3
A high-voltage probe can
be used with the DM 501 Digital Multimeter In lieu of the DC voltmeter. See the Tektronix Catalog tor a
list
ol DM 501 optional accessories.
'Two
identical
amplifiers are required to adjust x-y
phase difference.
'Requires
a
TM
500-series power module.
3-1
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......