Specification and Performance Check-5110
PERFORMANCE
CHECK
PROCEDURE
1.
Check
Trace Alignment
a.
Position
the
horizontal trace over the center
horizontal
graticule line.
b.
Check—that
the trace is
parallel to the graticule line.
c.
Adjust—
the TRACE
ROTATION control (rear-pane!
screwdriver
adjustment) to align
the trace
horizontally.
2.
Check
Geometry
a. Press
the
POWER switch to turn off the os
cilloscope.
b.
Interchange
the amplifier and time-base units in
their
respective
compartments.
Pull the
POWER switch to
on.
c.
Position
the
vertical trace over the center vertical
graticule line,
extending
vertically above and
below the
graticule
area, and
set the FOCUS and INTENSITY
controls
for a well-defined trace.
d.
Check—that
vertical
bowing and tilt of
the trace
display is
less than 0.1 division at the center line
and
when
positioned
horizontally across the entire graticule area.
e. Press the POWER
switch to turn off the oscilloscope
and
interchange
the amplifier
and time-base units back to
their usual compartments.
f.
Pull the POWER switch to on.
3.
Check
Beam
Finder
a.
Set
the
INTENSITY control for a dim trace.
b.
Press and hold
the BEAM FINDER
pushbutton
in,
then rotate the
position control of the vertical amplifier
and time-base
units
fully clockwise and counterclockwise.
c.
Check—
that
the display is intensified, compressed,
and
remains within the graticule area.
d.
Release the
BEAM FINDER pushbutton and return
the
INTENSITY control to
a normal setting.
4.
Check
Amplifier
Alternate
Operation
a. Push
both CH 1
and CH 2 pushbuttons
in and
position
the traces about two divisions apart.
b.
Set
the time-base unit Display
pushbutton to Alter
nate.
c.
Turn
the
time-base Seconds/Div
switch throughout
its
range.
d.
Check—
for trace alternation
at all sweep rates
(except
in amplifier
positions). At faster sweep rates,
alternation is
not
apparent; instead,
the display appears as
two
traces on the screen.
e. Press the POWER switch to tu rn off the oscilloscope
and
change
the amplifier from the vertical compartment to
the center
compartment.
f.
Pull
the
POWER switch on and repeat parts c and d
of
this step.
5.
Check
Amplifier Chop Operation
a.
Set
the time-base unit Display pushbutton to Chop.
b.
Turn the time-base Seconds/Div switch
throughout
its range.
c.
Check
—for a dual-trace
display at all sweep rates
(except in amplifier positions) without alternation.
d.
Press
the
POWER switch
to turn off the oscilloscope
and change
the amplifier
from the center compartment to
the
left vertical compartment.
e.
Pull the
POWER switch to on and repeat parts b
and
c of this
step.
6.
Check
Chop Operation Between Amplifiers
a.
Press the
POWER switch to turn off the os
cilloscope.
Install
a second vertical dual-trace
plug-ln unit
in
the center plug-in compartment
and set its controls for
dual-trace operation. Pull the POWER switch to on.
b. Turn the time-base Seconds/Div switch throughout
its
range.
c.
Check—for
two traces for each amplifier (one for
each channel) at all sweep
rates.
NOTE
If
it
single-channel amplifier
is
used instead of the
second
dual-trace
amplifier, the single-channel
trace will appear
once per sweep.
7.
Check
Alternate Operation Between Amplifiers
a.
Set
the
time-base Display
pushbutton to
Alternate
and
the
Seconds/Div switch
to 50
ms.
b. Check—for two traces for the left amplifier (one for
each
channel), then two traces
for the right amplifier,
alternately between
amplifier units.
2-7
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......