Signal D elay
Two standard UHF connectors are provided on
the rear of the instrument for insertion of an
external length of delay cable into the vertical-
input signal circuit. Choice of the appropriate
length and type of cable is at the discretion of
the user; no delay cable is furnished with the
instrument. A signal delay permits the sweep to
be triggered and under way before the signal is
applied to the vertical deflection plates.
SWEEP CIRCUIT
Type
Triggered, hard-tube bootstrap sweep circuit
with inverter to produce balanced deflection.
Rates
An eleven-position switch selects rates of .02,
.05, .1, .2, .5, 1, 2, 5, 10, 20 and 50 MICRO-
SECONDS/CM, with a maximum displacement
error of 2 percent over the center 8 cm of the
10-cm sweep length.
Sweep Starting Time
The horizontal sweep starts approximately
100-m^sec after the signal or triggering pulse
arrives at the rear-panel connector. An inserted
signal delay of approximately 150 m//,sec permits
the sweep to be triggered and under way before
the signal is applied to the vertical deflection
plates.
Duty Cycle Limitation
A duty-cycle limiting control automatically
limits the duty cycle of the sweep circuit to about
10 percent to avoid exceeding the dissipation
limits of some of the circuit components. The
limiting system serves purely a protective func
tion and does not provide a frequency dividing
operation.
The following table shows the maximum per
missible repetition rate for each of the avail
able sweep times per centimeter.
SWEEP TIME
MAXIMUM
REPETITION RATE
50 jasec/cm
600 c/s
20
/xsec
/cm
1.5 kc
10 jusec/cm
3 kc
5
use
c/cm
6 kc
2 jasec/cm
10 kc
1
fxse
c/cm
20 kc
.5 jitsec/cm
50 kc
.2
fxsec
/cm
50 kc
.1
f i
sec/cm
50 kc
.05
fis e c /
cm
50 kc
.02
fj
isec/cm
50 kc
Triggering
A triggering phase-inverter amplifier in con
junction with a selector switch permits the sweep
circuit to be triggered from either positive- or
negative-going portions of the observed signal,
or from positive or negative triggers from an
external source. A trigger voltage range of 100
volts to 3000 volts amplitude will be adequate
for stable triggering. The MARKER position on
the selector switch must be used when time
markers are desired.
Sweep Mode
A two-position switch provides for either re
petitive or single-sweep operation. When the
switch is in the single-sweep position, pressing
the RESET button arms the sweep circuit. The
sweep can then be triggered internally, by MAN
UAL TRIGGER, or by an external trigger. The
1-2
GENERAL DESCRIPTION — TYPE 5 0 7
Содержание 507
Страница 5: ...TYPE 507 OSCILLOSCOPE AND ACCESSORIES ...
Страница 6: ... I I I I I I I I I I I I I I I I I I I I I I I 1 1 I I I I ...
Страница 8: ... H I 1 ...
Страница 14: ......
Страница 20: ... t t t L L L L I L ...
Страница 27: ...V1I6 V 14 S VI53 V133 VI44 VI 4 r Li ...
Страница 28: ......
Страница 30: ......
Страница 32: ......
Страница 46: ...2 116 59 r D B TYPE 507 OSCILLOSCOPE A BLOCK DIAGRAM BLOCK DIAGRAM ...
Страница 48: ... 475V 1 TYPE 507 OSCILLOSCOPE SWEEP GENERATOR SWEEP GENERATOR ...
Страница 49: ... TYPE 507 OSCILLOSCOPE TIME MARK GENERATOR Z J3 TIME MARKER ...
Страница 50: ... 750 V TYPE 507 OSCILLOSCOPE A ...
Страница 51: ...HORIZONTAL AMPLIFIER 10 29 58 d O J S HORIZONTAL AMPLIFIER ...
Страница 54: ...T602 TYPE 507 OSCILLOSCOPE POWER SUPPLY ...
Страница 56: ...TYPE 507 OSCILLOSCOPE A CRT CIRCUIT ...
Страница 57: ......
Страница 59: ... 1 ...