![Tektronix 5 Series Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/tektronix/5-series/5-series_service-manual_1078777029.webp)
2. Carefully hold together the carrier interface assembly and turn it upside down so that the metal carrier tray is facing up. Support the
end of the carrier interface assembly away from the heat sink assembly so that it is level.
3. Carefully lift the metal tray from the assembly to expose the bottom of the carrier interface board and the M.2 SSD memory.
4. Remove the one T-8 Torx screw and remove the M.2 SSD memory board. Store or destroy the M.2 SSD memory board in accordance
with your organization's guidelines.
5. Carefully align and set the carrier board tray back onto the carrier interface assembly.
6. Carefully hold together the carrier interface assembly and tray and turn it so that the circuit board and heat sink assembly are facing
up.
7. Install the seven T-10 Torx carrier board screws. Tighten to 0.65 N·m.
8. Install the five T-8 Torx screws on the carrier board heat sink. Tighten to 0.45 N·m.
9. Reassemble the instrument and return it to the nearest Tektronix Service Center for replacement of the M.2 SSD board and
recalibration.
Troubleshooting 5 Series MSO
CAUTION: Before performing this or any other procedure in this manual, read the General Safety Summary and Service Safety
Summary found at the beginning of this manual.
To prevent possible injury to service personnel or damage to electrical components, please read information on Preventing ESD.
(See
This section contains information and procedures designed to help you isolate faults to a module.
This section requires that service personnel have the appropriate skills to work on this instrument, including PC troubleshooting and
Microsoft Windows operating system skills. Details of PC and Windows operation and service are not in this manual.
For assistance, contact your local Tektronix Service Center.
Maintenance
5 Series MSO Service Manual MSO54, MSO56, MSO58
29