![Tektronix 485 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/tektronix/485/485_instruction-manual_1078906008.webp)
Specification—485/R485 Service
DIFFERENCE Between Two Dial
Readings, Major D.T.P. Divisions
Fig. 1-3. Specified Differential Delay Accuracy vs. Size of Measure
ment.
Jitter
Less than 0.005% (one part in 20,000) of full scale plus
0.2 ns. (Wrap-around cover must be in place.)
X -Y
Full Sensitivity
(CH 1 Vert, CH 2 Horizontal.) 5 mV/div to 5 V/div in
10 calibrated steps (1-2-5 sequence) accurate within 2%.
Gain can be recalibrated at front panel.
Y Performance
Identical to CH 1. See Vertical Deflection.
X-Axis Bandwidth
DC to 4 MHz (with 10 div reference signal).
X-Y Phase Match (Full BW and BW Limit)
Within 3° to 4 MHz.
X-Gain Match to CH 2
+15°C to +35°C: 1%
—15°C to +55°C: 3%
CRT
T E K T R O N IX CRT
4-inch rectangular tube; 8X 10 divisions (0.8 cm/div)
display area. P31 phosphor normally supplied; P11
optional without extra charge. 21 kV accelerating
potential.
Photographic Writing Speed
A t least 3 div/ns for standard P31 phosphor; at least
6 div/ns
for optional
P11
phosphor
(with
the
TEKTRONIX C-31-R Camera and POLAROID 10,000
ASA film ).
Auto Focus
Automatically maintains beam focus for all intensity
settings. (Intensified zone and EXT Z axis are not
auto-focused).
Graticule
Internal, no parallax. Variable edge-lighting. Markings
for measurement of risetime. Graticule is dark with
illumination off.
Beam Finder
Limits display within graticule area.
External Z-Axis
Risetime 15 ns. Input R 500 T2. +0.2 V (DC to 20 MHz)
blanks trace of average intensity. +2 V (DC to 2 MHz)
blanks maximum intensity trace.
Beam Current Limit
Automatically limits the average beam current to protect
the CRT phosphor. (Limits average current to 20 pA for
sweep rates faster than 50 ms/div; 5 pA for 50 ms/div
and slower sweep rates and for X-Y. Backup system
shuts down power supply if average current reaches
approximately 30 pA.)
©
1-5
Содержание 485
Страница 3: ...485 R485 rviice T h 5 0 Oscil oscope ...
Страница 32: ...Maintenance 485 R 485 Service 4 4 Fig 4 1 Electrode configuration for semiconductors in this instrument ...
Страница 74: ...Rackmounting 485 R485 Service 6 6 ...
Страница 128: ...485 R485 Service MAR 1979 5 m V B A L ADJ ...
Страница 134: ... See Parts List for Fig 8 5 A 7 Partial Vertical am plifier board serial num ber ranges R E V D S E P T 1974 ...
Страница 137: ...Fig 8 6 A 7 Partial Vertical am plifier board See Pgrts Ljsf for serial number ranges f R E V B S E P T 1974 ...
Страница 145: ...Fig 8 9 A l l Partial Horizontal amplifier board REV E AUG 1975 See Parts List for serial number ranges ...
Страница 152: ... 0 iT s i ...
Страница 153: ...W H I T C X Y I6 J ROM Q S X J I 4 8 5 O S C U L O C O P E II93 4S REV D A J l 57fe HORIZONTAL A M P L IF IE R U ...
Страница 159: ...OSCILLOSCOPE 3 0 0 0 V TE ST P O IN T M AR K E D 2 9 5 0 V IN SOME INSTRUMENTS R EV e M AR 19 70 7 93 52 MRS ...
Страница 164: ...2S2 n eKn s e s 5 4 85 Os c illo sc o p e REV F A U S 1 9 7 7 I 1 9 3 5B POWER SUPPLY 773 ...
Страница 168: ... O I P 4 c N ifl 0 t 4 8 5 O S C I L L O S C O P E 2 d2 2 m u J tP Z O u u r 7 OJO d U O Hi 0u o o J ...
Страница 178: ......
Страница 179: ...485 R485 OSCILLOSCOPE REV D MAR 1977 ...
Страница 184: ...REV C MAR 1976 485 R485 OSCILLOSCOPE ...
Страница 197: ......