Maintenance
Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual
4- 3
Clean the On/Standby switch using a dampened cleaning towel. Do not spray or
wet the switch directly.
CAUTION.
Avoid the use of chemical cleaning agents, which might damage the
plastics used in this oscilloscope. Use only deionized water when cleaning the
menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as a
cleaner and rinse with deionized water. Before using any other type of cleaner,
consult your Tektronix Service Center or representative.
Lubrication.
There is no periodic lubrication required for this oscilloscope.
Inspect the outside of the oscilloscope for damage, wear, and missing parts,
using Table 4--1 as a guide. Immediately repair defects that could cause personal
injury or lead to further damage to the oscilloscope.
Table 4- 1: External inspection check list
Item
Inspect for
Repair action
Cabinet, front panel,
and cover
Cracks, scratches, deformations,
damaged hardware.
Repair or replace defective
module.
Front-panel knobs
Missing, damaged, or loose
knobs.
Repair or replace missing or
defective knobs.
Connectors
Broken shells, cracked insulation,
and deformed contacts. Dirt in
connectors.
Repair or replace defective
modules. Clear or wash out dirt.
Carrying handle, and
cabinet feet
Correct operation.
Repair or replace defective
module.
Accessories
Missing items or parts of items,
bent pins, broken or frayed
cables, and damaged
connectors.
Repair or replace damaged or
missing items, frayed cables, and
defective modules.
If the display is very dirty, moisten the wipe with distilled water or a 75%
isopropyl alcohol solution and gently rub the display surface. Avoid using excess
force or you may damage the plastic display surface.
CAUTION.
To prevent getting moisture inside the oscilloscope during external
cleaning, use only enough liquid to dampen the cloth or applicator.
Exterior Inspection
Содержание 4000 Series
Страница 8: ...Service Safety Summary vi Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 10: ...Environmental Considerations viii Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 13: ...Operating Information ...
Страница 14: ......
Страница 16: ...Operating Information 1 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 17: ...Theory of Operation ...
Страница 18: ......
Страница 22: ...Theory of Operation 2 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 23: ...Adjustment Procedures ...
Страница 24: ......
Страница 37: ...Maintenance ...
Страница 38: ......
Страница 55: ...Mechanical Parts List ...
Страница 56: ......