Checks and Adjustments
5. Using the approximately 1-kHz probe-adjustment square
wave as the input, obtain a 5-division display of the signal.
6
. Set the SEC/DIV switch to display several cycles of the signal.
Use the Channel 1 POSITION control to vertically center the
display.
Figure 4-1.
Probe compensation.
7. Check the waveform presentation for overshoot and rolloff
(see Figure 4-1). If necessary, perform step
8
to adjust the
probe low-frequency compensation. Otherwise, proceed to
step 9.
8
. Rotate the sleeve on the probe head to expose the LF COMP
adjustment capacitor (see Figure 4-2). Use a low-reactance
alignment tool to adjust LF COMP and obtain a square, or
nearly square, front corner on the waveform tops (Figure 4-1,
top trace).
9. Select CH 2 VERTICAL MODE. Disconnect the Channel 1
probe tip and attach the Channel 2 probe tip to the PROBE
ADJUST output terminal.
10. Obtain a 5-division signal display and vertically center the
trace.
11. Repeat step 7 for the Channel
2
probe.
4-2
2225 Operators
Содержание 2225
Страница 12: ......
Страница 13: ...SECTION 1 PREPARATION FORUSE 2225 Operators ...
Страница 14: ...1 ...
Страница 20: ......
Страница 21: ...SECTION 2 OPERATION 2225 Operators ...
Страница 23: ...NOTE Numbers on the instrument are keyed to Table 2 1 ...
Страница 36: ...SECTION 3 APPLICATIONS 3 2225 Operators ...
Страница 37: ...3 ...
Страница 54: ...SECTION 4 CHECKSAND ADJUSTMENTS 4 2225 Operators ...
Страница 55: ......
Страница 58: ...Checks and Adjustments Figure 4 2 Locating the low frequency compensation adjustment 2225 Operators 4 3 ...
Страница 59: ......
Страница 60: ...SECTION 5 PERFORMANCE CHARACTERISTICS 5 2225 Operators ...
Страница 61: ...5 ...
Страница 69: ......
Страница 70: ...SECTION 6 OPTIONSAND ACCESSORIES 6 2225 Operators ...
Страница 71: ...6 ...
Страница 77: ......
Страница 79: ...NOTES ...