Calibration—221 Service
5. Check Battery Charger Operation
a. Connect the common lead of the digital voltmeter to
the second pin connector on the power supply board.
Connect the other lead of the digital voltmeter to the BC
test point. See Fig. 5-2.
b. Vary the line voltage from 90 to 250 Vac w ith the
instrument on or off.
c. Check fo r a reading of +0.6 to +0.8 o f a vo lt on the
digital voltmeter.
d. Check fo r a 0 volt reading on the digital voltmeter,
when the instrument is disconnected from the power line.
+75 Volts 5th Pin Connector From Rear
+48 Volts 4th Pin Connector From Rear
OANC
i
2nd Pin Connector From Rear
Battery Charge (BC)
Test Point
Fig. 5-2. Location of the +75 volt, +48 volt and the battery charger
test points.
(Behind Pot)
TP
R662
High Voltage
- 1 0 0 0 V
TP
R644
Low Voltage
0 V TP
Fig. 5-1. Location of the Low Voltage, High Voltage adjustments
and test points.
REV. APR 1974
5-5