Calibration-Type
1 84
a. Positive Amplified Markers
b. Negative Amplified Markers
Fig. 6-1 4. Marker Amplifier Otuput Signals into
1
kr!. Volts/ em
1 0,
Time/em (Table
5-3 1 .
d. Change the MARKER AMPLIFIER switch to the (-)
position and the test oscilloscope Trigger Slope switch to
(-Int) position.
e. Check the ampl itude of the Marker Amplifier output
signal for a minimum -25 volt signal. See Fig. 6-1 4b.
f. Remove the adapters and cables from the Type 1 84
and test oscil loscope.
TABLE
6-3
Type
1 84
Test Osci l l oscope
MARKER SE LECTOR
Time/Cm
1 p.S
1 0 p.SEC
5 p.S
10 p.SEC
1 0 p.S
.1 mSEC
50 p.S
.1 mSEC
.1 mS
1 mSEC
.5 mS
1 mSEC
1 mS
10 mSEC
5 mS
10 mSEC
1 0 mS
1 00 mSEC
50 mS
1 00 mSEC
.1 S
1 SEC
.5 S
1
SEC
1 S
1 SEC
5 S
1 SEC
NOTES
6-1 4
®