![Tektronix 070-7169-05 Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/tektronix/070-7169-05/070-7169-05_user-manual_1077884097.webp)
Appendix B: Operator Performance Checks
B–8
1502C MTDR User Manual
If the aberrations are out of specification, the ohms-at-cursor function might be
less accurate than specified.
1. Connect the 50
W
precision terminator to the front-panel CABLE connector.
2. Set the DIST/DIV control to 5 ft/div (1 m/div).
3. Increase the VERT SCALE control to 50 m
r
/div.
4. Using the
n
o
POSITION control, move the top of the pulse to the center
graticule line.
O
F
F
O
F
F
O
F
F
O
N
ac
–1.872 ft
Figure B–10: Top of Pulse on Center Graticule
5. Set the DIST/DIV control to 0.2 ft/div (0.05 m/div).
6. Turn the
n
o
POSITION control clockwise until the rising edge of the
incident pulse is in the left-most major division on the display.
O
F
F
O
F
F
O
F
F
O
N
ac
1.744 ft
Figure B–11: Rising Edge of Incident Pulse in Left-most Major Division
7. Using the
n
o
POSITION control, move the cursor back to 0.000 ft (0.00 m).
6. Aberrations Check
Содержание 070-7169-05
Страница 15: ...General Information x 1502C MTDR User Manual...
Страница 19: ...General Safety Summary xiv 1502C MTDR User Manual...
Страница 79: ...Operator Tutorial 2 28 1502C MTDR User Manual...
Страница 83: ...Options and Accessories 3 4 1502C MTDR User Manual...
Страница 89: ...Appendix A Specifications A 6 1502C MTDR User Manual...
Страница 117: ...Appendix D Application Note D 10 1502C MTDR User Manual...
Страница 127: ...Index Index 4 1502C MTDR User Manual...