Grade 2 of password (set by manufacturer as 03210):
Users can change parameters codes 1 to 25.
•
Grade 3 of password (set by manufacturer as 06108):
Users can change parameters codes 1 to 26 and 35 to 37.
•
Grade 4 of password (Please contact Tek-Trol for this password if you need):
Users can change parameters codes 1 to 30 and 35 to 37.
•
Grade 5 of password (Fixed):
Users can change parameters codes 1 to 46.
NOTE
Password Grade 5 can be set by skilled users. Grade 4 is mainly used for resetting
total volume in password. Grades 1 to 3 can be set by anyone.
There are 46 parameters of Tek-Flux 1400A, user can set every parameter. The list of
parameters is as shown below:
Code
Parameter
Setting
Grades
Range
1
Language
Select
2
English
2
Comm. Address Set Count
2
0 to 99
3
Baud Rate
Select
2
600 to 14400
4
Sensor Size
Select
2
5
Flow Unit
Select
2
l/h, l/m, l/s, m
3
/h, m
3
/m, m
3
/s
6
Flow Range
Set Count
2
0 to 99999
7
Flow Rspns
Select
2
1 to 50
8
Filt Factor
Select
2
0/10/15/30/60/90/120/150/180/240
9
Flow Direct
Select
2
Plus/ Reverse
10 Flow Zero
Set Count
2
0 to ±9999
11 Flow Cutoff
Set Count
2
0 to 599.99%
12 Cutoff Ena
Select
2
Enable/Disable
13 Total Unit
Select
2
0.001m
3
to 1m
3
, 0.001l to 1l
14 SegmaN Ena
Select
2
Enable/Disable
15 Analog Type
Select
2
0 to 10mA /4 to 20mA
16 Pulse Type
Select
2
Freque / Pulse
17 Pulse Fact
Select
2
0.001m
3
to 1m
3
, 0.001l to 1l
18 Freque Max
Select
2
1 to 5999 HZ
19 Mtsnsr Ena
Select
2
Enable/Disable
20 Mtsnsr Trip
Set Count
2
59999 %
21 Alm Hi Ena
Select
2
Enable/Disable
22 Alm Hi Val
Set Count
2
000.0 to 599.99 %
23 Alm Lo Ena
Select
2
Enable/Disable
24 Alm Lo Val
Set Count
2
000.0 to 599.99 %
25 Sys Alm Ena
Select
2
Enable/Disable