6.Troubleshooting
N2SERIES
Symptom
Cause
Remedy
Optics
The nosepiece is not in the right
position.
Turn the nosepiece into
the right position.
(1)The side of the
field of view is dark
or not even.
Stain or dust has accumulated on
the condenser, objective,
eyepieces, base lens. .
Clean the lens.
Stains have accumulated on the
specimen.
Clean the specimen.
(2)Stain or dust is
observed in the field
of view.
Stains have accumulated on the
lens.
Clean the lens.
No cover glass on the specimen
slide.
Add the cover glass.
The cover glass is not standard.
Use a standard cover glass
with thickness 0.17mm.
The specimen faces down.
Make it face up.
The immersion oil has
accumulated on the dry objective.
Clean thoroughly.
The immersion oil is not used for
oil objective 100XR.
Use immersion oil.
Air bubble in the immersion.
Get rid of the air bubble.
Use wrong immersion oil.
Use a correct one.
The aperture is not opened to
correct size.
Adjust the iris diaphragm.
Stain or dust has accumulated on
the lens in the inlet of the head.
Clean the lens.
(3)Unclear image
The condenser is not in the right
position.
Adjust the condenser.
The specimen slide is not fixed.
Fix with clips.
(4)One side of the
field of view is dark
or the image moves
while focusing.
The nosepiece is not in the right
position.
Turn the nosepiece into
the right position.
The iris diaphragm is not big
enough.
Adjust the iris diaphragm.
The condenser is not in the right
position.
Adjust the condenser.
(5)The field of view
is not bright enough.
Stain or dust has accumulated on
the condenser, objective,
eyepieces, base lens.
Clean the lens.