4) Press “
∆
” button, you can use the relative function to eliminate the zero error.
5) Connect the probes across capacitor to be tested.
Note: When testing 400
μ
F / 4000
μ
F capacitor, note that there will be approx. 30 sec time lag.
Caution: Ensure that the circuit to be tested is "dead".
Max. input over-load: 250V RMS. and<10sec.
2.8 Frequency measurement
1) Connect the black test lead to "
COM
" socket and red test leads to the "
FREQ
" socket.
2) Set the selector switch to “
DUTY
FREQ
” position.
3) Connect the probes across the source or load under measurement.
2.9 Temperature measurement
1)
Set the selector switch to the ”
TEMP °C/°F
” range and connect the k type thermocouple’s
black ( or “-“ terminal ) test lead to “
COM
” socket and red ( or “+“ terminal ) test lead to the
"
TEMP
" socket and read value on display screen. The range of TP-02 thermocouple is
-50~+700°C ( -58 ~1292°F).
2)
Press
SELECT
button to switch °C or °F display.
3)
Opening the input will display the temperature at the meter terminals by the internal built-in
sensor.
2.10 Auto Power Off
When the meter has been turned on about 15 minutes without any action from the users, the
meter will automatically change to “OFF” mode, and you can press any function button to
wake up again.
If press and hold
SELECT
button to turn on the meter, the AUTO POWER
OFF function will be canceled.
2.11 Using Communications
1) Connecting the RS232 cable to the meter and a computer as shown in Figure 1, then turn on
the meter.
2) Press button more than 2 seconds on the panel to start the RS232 communication
function. If you want to cancel the RS232 function, please press the button more than
2 seconds once again.
3) About the father information of communication, please refer to the on-line help of Data Logger
attached.
3.CARE AND MAINTENANCE