Teilch TCH PRO 203 Скачать руководство пользователя страница 2

Screen Symbols 

SD Inserted 

SD Busy - Do Not Remove 

Data Saved to SD 

Data Not Saved (SD not present or full) 

Save 

Lock. Default password is 1234. To 
reset password, press reset button.  

Settings 

         Navigation 

Red Dot: Coincidence Loss Error 

teilch

micro-contamination and optical instrumentation

TCH PRO 203 -

 

QUICK START GUIDE

 

OPERATIONS

 

To start using the instrument, proceed as follows: 

1. Position the instrument in the environment 

to be measured. 

2. Remove the 

red

 protective cap from the 

inlet tube. 

3. Turn the device 

ON

 by pressing the Power 

button and holding it for 3 seconds. 

4. After pressing the power key, a SPLASH 

screen will appear for three seconds, 
displaying the company logo, model 
number, serial number, and firmware 
version number. The Start screen displays 
on the LCD. 

5. When the MENU screen appears, the 

instrument is ready for use. Select mode 
and follow on-screen instructions. 

Cleanroom Particle Analysis: 
Cleanroom certification and 
monitoring 

Continuous Particle Log: Save full 
particle log data into SD Card 

Air Quality Analysis: Particulate Matter  

Pressure: Differential Pressure 
Sampling 

To purge sensor, install Inlet Filter and run any 
Count cycle (pump ON) for 1 minute 

ADVANCED OPERATIONS 

Data Management and Analysis: 

Data may be 

viewed and analyzed on instrument screen or 
may be saved into SD Card.  All data is 
formatted as text and may be viewed and 
analyzed with any text or spreadsheet software 
(not provided by Teilch).  

Firmware Upgrade: 

Press Reset while holding 

Power button. Follow on screen instructions. 

Touch Screen Calibration: 

If screen touch panel 

does not work accurately, turn unit off, press 
and hold Count and press Power while touching 
screen. Follow on screen instructions.

Laser Safety:

 Class I laser-based device. 

No exposure to laser radiation occurs 
during normal operation.  

DO NOT 

remove any housings or covers.

Important: 

There are no user-serviceable 

parts inside this instrument. All repair and 
maintenance needs should be done by a 
qualified technician.  