![Teilch TCH Pro 202 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/teilch/tch-pro-202/tch-pro-202_operation-manual_1075516021.webp)
Operation Manual TCH Pro 202 & 203
2. Continuous Particle Log
Purpose
: To log particle counts with 10nm resolution for up to 20,000 particles.
CLEANROOM PARTICLE ANALYSIS > ALARMS
Set up independent alarms for absolute counts and
maximum concentration.
•
Size
: Select your particle size or leave it on disabled.
•
Counts Alarm
: Set particle size and maximum allowed
total particle number.
•
Concentration Alarm
: Set particle size and maximum
allowed particle concentration.
CLEANROOM PARTICLE ANALYSIS > OTHER
SETTINGS
Set up storage settings.
•
Auto
: Automatically saves data.
•
Manual
: Device will ask after sample set/count routine if
user wants to save data.
•
Save Displayed Data
: Save a summary of 8 common
channels displayed in first screen of particle analysis.
•
Save Full Data
: Save data for sixty particle size bins. Bins
are organized in logarithmic type array.
•
Between 0.2
μ
m and 1
μ
m, particle sizes are grouped in
20nm size bins.
•
Between 1
μ
m and 3
μ
m, particle sizes are grouped in
250nm size bins.
•
Between 3
μ
m and 7
μ
m, particle sizes are grouped in
0.5
μ
m size bins.
•
Between 7
μ
m and 10
μ
m, particle sizes are grouped in
1
μ
m size bins.
MENU
SCREEN
Page of
21
33