Connect the opposite end of the Thermocouple Cable (or the single
-
connector end
of the Split Thermocouple Cable) to the Thermocouple Calibrator thermocouple
output.
The UUT display will indicate as follows:
a.
Line 1: CALA
b.
Line 2: temperature in °C
c.
Line 3: temperature in tenths of °C (out to 0.000001 °C)
23.
Set the Thermocouple Calibrator
thermocouple type
to Type
E
24.
Set the calibrator to output 0.0 °C.
25.
Allow at least
five minutes for this connection to stabilize.
26.
Press UUT
to set
the UUT display equal to the
Thermocouple Cable calibrated
correction value ± 0.02 °C.
27.
Press UUT
.
28.
The display will change to rESA showing the actual measured value saved in the
previous steps.
29.
For single channel
UUTs
, skip
to
Step 35
below
.
30.
For two channel
UUTs
, continue with Step
a.
For two channel UUTs using the single
-
ended Thermocouple Cable,
disconnect the cable from the UUT Channel 1 input, and reconnect the
cable to the Channel 2 input. Repeat Step 25
31.
Press UUT
.
32.
The UUT display will indicate as follows:
a.
Line 1: CALb
b.
Line 2: temperature in °C
c.
Line 3: temperature in tenths of °C (out to 0.000001 °C)
33.
Repeat Steps 26
and 27
34.
The display will change to rESb showing the actual measured value saved in the
previous steps.
35.
Press UUT
to save the current alignment values and return the UUT to normal
operation.
5
If substituting another thermocouple type, set the Thermocouple Calibrator as appropriate for
the selected thermocouple type.