Operating Instructions
3-
4
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
4.
To save the current
parameter value
and cycle
to the next parameter,
press
5.
To save the current
parameter value
and exit
the Setup Menu, press
;
6.
To disregard changes
made to the current
parameter
value
and exit
the Setup Menu, press
.
If no key is pressed for
10
seconds, the instrument will save the current configuration
and exit the Setup Menu.
View Modes and Statistics
The instrument
features multiple view
modes including
a variety of real
-time statistics
, all
available
at the touch of a button.
Figure 6
below describes each view
mode.
Press
to
change view modes
.
For each mode, the active measurement or statistic result
is
displayed on the second line of the display.
P
ARAMETER
AVAILABLE VALUES
Thermocouple Type
E, J, K, T
Temperature Units
°C, °F, K
T1 Probe Offset
±0.1 ° increments
T2 Probe Offset
1
±0.1 ° increments
Open Wire Detection
ON, OFF
Resolution and
Rounding functions
rndG, Auto, CEIL, nOrL,
trnC
trnd
ON, OFF
1
T2 Probe Offset available
on model
922B
only.
Figure 5: Setup Menu Parameters and Values
V
IEW
M
ODE
D
ISPLAY
I
NDICATOR
D
ESCRIPTION
T1–T2
T1-T2
Current Channel 1 measurement
–
current
Channel 2 measurement
Minimum
MIN
Minimum temperature recorded during
current session
Maximum
MAX
Maximum temperature recorded during
current session
Average
AVG
Average of all temperatures recorded during
current session
Range
RNG
Maximum
-
Minimum
Standard
Deviation
STDEV
Standard deviation of all temperatures
recorded during the current session
1
.
1
Standard Deviation is calculated using the population formula:
𝜎𝜎
=
�
∑
(
𝑥𝑥−
µ)
2
𝑁𝑁
Figure 6: View Modes and Statistics