Preparation for Use
2-8
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
connection to the instrument.
3.
The instrument
will immediately display a temperature measurement
for the
connected channels. However, to ensure
valid and best accuracy measurements,
continue to Step 4
below;
4.
Set the desired measurement parameters as follows:
a.
Enter the Setup Menu by pressing
, hold the key down for
approximately 1.5 seconds, and then release
it;
b.
The active thermocouple
t
ype is flashing on the
display
. Use
to
select
the
thermocouple type
of the connected temperature probe
(E, J,
K, T, B, N, R, or S);
c.
Momentarily (do not hold) press
to save your selection
and move to
the next parameter
;
d.
The active temperature unit is flashing on the display
. Use
to
select the desired temperature unit (°C, °F, or K);
e.
Momentarily press
to save your selection and move to the next
parameter;
f.
Channel 1 probe offset value is flashing on the display. If the
temperature probe’s offset value is known, press
to set the
Channel 1 probe offset to the probe’s offset value. See
Section 3.10,
for more information
.
g.
Momentarily press
to save your selection and move to Channel 2
probe offset (if equipped);
h.
If desired, repeat Step (f) above for Channel 2;
i.
Momentarily press
to save your selection and exit the setup
menu
.
Congratulations! You’r
e
now ready to make accurate and reliable temperature
measurements, wherever and whenever you may need to.