![Tegam 138-415-1 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/tegam/138-415-1/138-415-1_instruction-manual_1075395014.webp)
Operation & Service Manual
Section I
Model 1805B RF Control Unit
Introduction and General Description
Table 1-2 Recommended Consumable Materials
Table 1-3 Special Tools and Test Equipment
1-7
1
Acid Brush (Fiber Bristle)
H-B-643 Type II CLASS
I
2
Aluminum Wool
MIL-A-4864A
3
Brush, Soft Bristle
H-B-420 Type II
4
Cleaner/Solvent
Isopropyl Alcohol TT-1-735A (3)
5
Cloth,Lint Free
MIL-C-85043 Type IIA
6
Filter Coat Adhesive
7
Permanent Polymer Coating IPC-SM-840 Class III
(Solder Mask)
TOOL/EQUIPMENT/ NOMENCLATURE USE AND APPLICATION
MODEL NUMBER
Gigatronics GT9000 or Synthesized Signal Generator, 0.01-18 Operational Checkout/Calibration
equivalent GHz+10 dBm
HP3458 or equivalent Digital Voltmeter, 8 1/2 digits Operational Checkout/Calibration
HP436/437/438 or Power Meter
Operational Checkout/Calibration
equivalent
HP8481 or equivalent Power Sensor, Type N, 0.01-18 GHz Operational Checkout/Calibration
Weinschel 44-10 or 10 dB Fixed Attenuator, Type N, dc-
Operational Checkout/Calibration
equivalent
18 GHz
Weinschel 56-10 or 10 dB Fixed Attenuator, 3.5mm, dc-
Operational Checkout/Calibration
equivalent 26.5 GHz
SYSIIA-SureCal System IIA Windows Software
Operational Checkout/Calibration
(SureCAL)
Any
WindowsTM Compatible Printer
Operational Checkout/Calibration
with GPIB interface card and
Windows
TM 3.1 Operating System
Any
Windows
TM 3.1 Compatible Printer
Calibration Report/Graph/Test
Report Printout
Weinschel 1915 or Torque Wrench, 14±1 inch pounds
Connector Coupling Torque Type
equivalent (Type N)
N Connectors
Weinschel 080-502 or Torque Wrench, 7±1 inch pounds Connector Coupling Torque
equivalent (SMA, 3.5mm)
SMA. 3.5 mm Connectors
TEGAM 138-417 AC Extender Board Calibration Adjustment
TEGAM 138-417 AC Extender Board Calibration Adjustment