The OPT-1020 and OPT-1120 daughter boards provide the analog input signal
isolation for the DAQ cards.
OPT-1020

bandwidth 20kHz, isolation voltage 2.5 kV, screw terminals

OPT-1020B

bandwidth 20kHz, isolation voltage 2.5 kV, BNC connector

OPT-1120

bandwidth 120kHz, isolation voltage 1.5 kV, screw terminals

OPT-1120B

bandwidth 120kHz, isolation voltage 1.5 kV, BNC connector

If the DAQ card provides the 12V power supply voltages, the OPT-1020/1120 board
can be powered from the card; if not, external power supply must be used.
One OPT-1x20 board can be connected to the DAQ card using the CAB-2604C flat
cable and CAB-2511/2 shielded cable. If multiple OPT-1x20 boards are used, an
interconnection cables CAB-1004C and CAB-2604C (up to four boards), or
CAB-1008C and CAB-2608C (up to eight boards) are required (not included).
Entire set of boards can be connected to the DAQ cards via CAB-2511/2. Please pay
attention to the maximum current capacity of the DAQ cards power supply output,
for higher number of OPT-1x20 boards it is necessary to use external power supply.

General instructions for use

The OPT-1x20 boards is designed for DAQ&C applications and may be used only
according to the manufacturer's recommendations and precautions given in this
manual and other general standards and terms and may be used only such a way,
that its failure caused by any reason will not be dangerous to any person or property.

Installation

The OPT-1x20 boards is supplied as an unencapsulated kit intended for fastening
via three screws. 
The board can be used in an environment with operating temperature -10~60 °C and
relative humidity up to 90%, noncondensing and normal levels of pollution.

Specifications

Input voltage (for processing):

±11 V max.

Overvoltage protection:

±24 V

Input impedance:

1 MOhm

(±1 %)

Selectable gain:

1x, 10x

(via on-board jumper)

Bandwidth (-3 dB, signal 7 V

AC

):

0÷20 kHz typ.

(all OPT-1020 version)

0÷120 kHz typ.

(all OPT-1120 version)

Nonlinearity:

±0,05% max.

(±0,025 typ.)

Isolation voltage (see note below):

2.5 kV

AC

 / 3.5 kV

DC

(OPT-1020H only)

1.5 kV

AC

 / 2 kV

DC

   (all other version)

Power supply:

+10.8~12.6 V

DC

 

(240 mA max.)

Recommended cable length:

2 m max.

Dimensions of board:

102 x 49 mm 

Mounting hole spacing:

94 x 42 mm

Mounting hole diameter:

3.5 mm

Note:

"AC" indicates the rms value of a 50 Hz AC harmonic signal.

Configuration jumpers

JP1

jumper for input range configuration, resp. gain of input amplifier

JP2

jumper for configuration of OPT-1x20 output signal routing to the DAQ card
input (only one OPT-1x20 board can be connected to the card input)

JP3/4

jumpers for selecting of power supply (see picture above)

P1

potentiometer for precise gain adjustment 

P2

potentiometer for precise offset adjustment 

Pin assigment of connector PFL26 (J3), resp. cable CAB-2511 to the DAQ card
PFL26

D-Sub 25

signal

PFL26

D-Sub 25

signal

- - -

AGND

13 

AGND

3 25  AIN0

4 12  AGND

5 24  AIN1

6 11  AGND

7 23  AIN2

8 10  AGND

9 22  AIN3

10 

AGND

11 21 

AIN4

12 8 

AGND

13 20 

AIN5

14 7 

AGND

15 19 

AIN6

16 6 

AGND

17 18 

AIN7

18 5 

AGND

19 

17 

- - -

20 

- - -

21 

16 

- - -

22 

- - -

23 

15 

GND (power supply) 24 

GND (power supply)

25 

14 

+12V (from card)

26 

reserved

Pin assigment of connector PFL10 (J4)
PFL10

signal

1, 2, 3, 4

power supply 12V (+5%, -10%), positive signal

5, 6

unused 

7, 8, 9, 10

power supply 12V (+5%, -10%), negative signal

Jumpers JP3/JP4 must be removed before connecting an external power supply.

JP1

AIN7 . . . . AIN0

J3

J1

J4

JP2

JP3/JP4

gain 1x

P1

P2

+IN
-IN

pin 1

pin 1

gain 10x

power supply from DAQ card 

power supply from an external
source via J4 connector

DAQ card input selection 