39
1.Remove the
battery compartment
2.Remove the
case bottom
3.Install fuses with
correct type and
specification
Figure 5-1. Fuse Replacement
B. Replacing the Fuses
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
Warning
Follow Figure 5-1 and proceed as follows to replace
the Meter’s fuse:
Turn the rotary switch to
OFF
and remove all
connections from the terminals.
Remove the screw from the battery compartment,
and separate the battery compartment from the case
bottom.
Remove the three screws from the case bottom,
and separate the case bottom from the case top.
Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
Install ONLY replacement fuses with the identical
type and specification as follows and make sure the
fuse is fixed firmly in the bracket.
Fuse 1: F0.5AH/250V, ø5 x 20mm
Fuse 2: F10AH/250V, ø5 x 20mm
Rejoin the case bottom and case top, battery
compartment and case bottom, and install the 5
screws.
Replacement of the fuses is seldom required. Burning
of a fuse always results from improper operation.