4
In case of two part system, install the fork probe on vessel, connect wires/ cable between fork probe
and controller, refer fig 2 .
4.
Maintenance
If the material has built up tendency, over a period of time, tines should be cleaned whenever need occurs.
Ensure that the cable glands and the housing lid are sealed to prevent ingress of moisture.
5.
Trouble Shooting
SL Problem
Cause
Solution
1. Switch not working 1.
Loose connection or disconnected
power supply connection
2.
Incorrect power supply
3.
Wrong probe connections in case
of two part system
4.
Forks are not vibrating due to
bend
1.
Check and tighten connection
2.
Check and connect correct power
supply. Ensure correct polarity for
24 VDC supply
3.
Check and make correct connection
of fork probe and controller.
4.
Check and install the fork probe
properly. Ensure forks are not bent.
Else contact service dept.
MAN/ VFSL/ REV 00/07‐22