3.5.     Detection walkthrough 

 
Plug the supplied USB cable into the device and then into a USB2.0 port of your computer.  

! Please DO NOT plug the USB cable in a USB 3.0 port ! 

(unless there is no 2.0 port present)

 

 
Run the Recognoil program (Recognoil 2.2.1) 
 
Prepare the object and put the Recognoil device facing down on the surface. 
 

1.

 

switch on the device 

The device will be switched on automatically upon correct USB connection. 

 

2.

 

detector activation 

If the detector serial number is correct, you can now choose one of two detection modes: 

 

3.

 

LIVE preview 

(only for Professional Edition)

 

Click  the  “LIVE”  button  (or  hold  the  trigger  1.5  to  4  secs.)  for  a  real-time  video  feed  of  the  irradiated 
surface.  Put  the  device  on  the  surface.  Press  and  hold  the  trigger  to  begin.  Exposure  time  is 
automatically selected. The result values are displayed in the bar graph diagram on the right hand side. 
The mode will be quit after touching the “LIVE” button again. If the device is detecting big amount of 
light entering the detector, it will stop working for the reason of personnel safety – risk of eye damage 
when viewed directly in to the UV source. 

 

4.

 

choosing the surface contamination 

This step is optional. The oil files are loaded in the /Recognoil/OilDB/ directory. Choose “LOAD OILS” to 
show the list. After the oil has been chosen from the list, the surface concentration and layer thickness 
values are shown. 

 

 

Warning  –  the  resulting  values  of  surface 
concentration  and  layer  thickness  are  strongly 
dependent  on  the  substrate  layer.  Please  be 
careful when interpreting the results. 

 
 

5.

 

Activating Photo mode 

Click on the PHOTO button or touch the trigger (0.2 to 1 sec). Put the device onto the surface and press 
the trigger. Please do not move the device during the scanning procedure (3-5 seconds). 

 

6.

 

Standard mode 

The standard mode offers automatic intelligent sequence detection. 

 

7.

 

ALD mode – ambient light deduction 

In case of improper light shielding, the detector switches itself in to the ALD mode. The detection now 
takes double the time, but the ambient light is eliminated. This function has its own limit of the amount 
of light that can be deducted. When the limit is exceeded, you will be noticed that the exposition could 
not be finished. 

 

8.

 

Manual mode 

In case of the automatic mode is not applicable, i.e. in need of very long or very short exposure times, 
you can click the MAN button and manually set the exposure time. This function also finds out, whether 
to use intelligent light deduction or not. 

Содержание RECOGNOIL

Страница 1: ...nual Software version 2 2 1_2015 02 21 1 RECOGNOIL Oil Layer Detector TechTest s r o Na Studankach 782 CZ 551 01 Jaromer http www techtest eu info techtest cz Recognoil oil layer detector User s manua...

Страница 2: ...t is not suitable for very refined mineral oils these applications often require a fluorescence agent The Recognoil device is not suitable for use with silicone oil based compounds Recognoil integrate...

Страница 3: ...ctivate the PHOTO mode This mode requires you to hold the device on one place 3 to 5 seconds after touching the trigger again After that a double beep is played marking the scanning complete 5 Please...

Страница 4: ...3 3 Software 7 3 4 SW Recognoil Advanced Professional 13 3 5 Detection walkthrough 14 3 6 Settings and Oil Pick dialog 16 3 7 Load RRW function 17 3 8 Battery charging 18 3 9 Notes on device operation...

Страница 5: ...erated only by an instructed person equipped with UV goggles No other person is allowed to share the same working place or must be also equipped with UV goggles 1 2 Workplace safety measures a Keep th...

Страница 6: ...l and dust contaminations degrade future measurement Do not use organic solvents when clearing Do not use paper towels for device cleaning The paper fibers show a great amount of fluorescence and thus...

Страница 7: ...ecognoil After that if battery voltage is O K you will be able to choose a detection mode After the detector activation choosing a mode and exposition trigger press the surface is irradiated and the r...

Страница 8: ...and choosing Run as Administrator This of course requires administrator privileges please contact your local network administrator in case of any problems Insert the supplied USB Flash disc in a free...

Страница 9: ...ftware version 2 2 1_2015 02 21 9 img 3 In the next Windows please select Custom and click Next img 4 The first item is to be checked along with uEye ActiveX control The other items need not be instal...

Страница 10: ...tall and wait a few seconds for the installation to be finished Now in the Additional features dialog there is no item necessary Please uncheck all fields and click Next After the installation has bee...

Страница 11: ...un After the first run of the software the default directory is c recognoil The result files are also stored in this directory You can change the path in Settings dialog Please do not choose any netwo...

Страница 12: ...ated with every Recognoil device After device connection and after the trigger has been pressed the map of detector output intensity DO is shown together with its arithmetic mean value of the whole ev...

Страница 13: ...Recognoil User s Manual Software version 2 2 1_2015 02 21 13 3 4 SW Recognoil Advanced Professional img 10 Software Recognoil Advanced Professional MENU...

Страница 14: ...o the UV source 4 choosing the surface contamination This step is optional The oil files are loaded in the Recognoil OilDB directory Choose LOAD OILS to show the list After the oil has been chosen fro...

Страница 15: ...the display or the mouse button The outcome is then computed only for this selected area The contrast slider may be locked to an exact value e g when performing scans of similar samples to see the co...

Страница 16: ...ome may not be the same as in the automatic intelligent mode For advanced users only You can disable the button sound effects by un checking Button sound effects Save RAW images the default setting is...

Страница 17: ...ng dialog the raw RRW Recognoil RAW files are saved in the working directory under Raw You can open these files later after the detection and evaluate and save it without the need of repeated surface...

Страница 18: ...5 you will not be able to continue working with the device Please leave the device plugged in a USB port or in a USB charger and let charge for minimum of 4 hours The full recharge takes up to 20 hou...

Страница 19: ...ct should be handed over to an authorized collection site for recycling waste electrical and electronic equipment EEE Improper handling of this type of waste could have a possible negative impact on t...

Страница 20: ...do not store in a damp environment do not store in temperatures below 5 deg Celsius and not exceeding 40 degrees when cleaning the device use a damp cloth do not use paper towels or organic solvents...

Страница 21: ...s the outcomes of the fluorescence process which emit radiation in the domain of visible light If there is a light coming from the environment not originating from the process of fluorescence it inter...

Страница 22: ...ensuring that no light can get inside the detector head If this rubber profile has been contaminated by touching a highly polluted surface it will hand over the residues to a surface which is cleaner...

Страница 23: ...nderlying surface Again a black anodic oxidized aluminum metal sheet is recommended for these purposes It is recommended to place the wire in a diagonal alignment to minimize the effects of unwanted U...

Страница 24: ...tate the material goes into the degreasing line and then would check the output The operator has already the knowledge how much DO detector output shows the Recognoil device after degreasing at a norm...

Страница 25: ...allowing you to perform validation test periodically Validation procedure can be executed from the Settings menu Please follow the instructions of every step In the final step the resulting value is s...

Страница 26: ...ration process takes up to 60 second to complete During this time you are obliged to hold the device in a steady position without any movement 4 After the calibration process has finished the calibrat...

Страница 27: ...ystem Windows XP 7 8 Recognoil Basic free of license Recognoil Advanced Professional for extra charge USB 2 0 512 x 768 px Approx 40 x 30 mm approx 18 x 12 mm 9V DC 365 nm 900 mW 15 Lithium Ion 3 7 V...

Страница 28: ...stallation interrupts with an error A Please run the driver pack as administrator i e by right clicking on the file and choosing Run as Administrator In case of any problem with privileges please cont...

Страница 29: ...Recognoil User s Manual Software version 2 2 1_2015 02 21 29 9 APPENDIX...

Отзывы: