Proline Promass F 500 EtherNet/IP
Diagnostics and troubleshooting
237
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
943 API pressure out of specification
1. Check process pressure with
selected API commodity group
2. Check API related parameters
0x800037F
Status signal
S
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
944 Monitoring failed
Check process conditions for
Heartbeat Monitoring
0x80001C6
Status signal
S
Diagnostic behavior [from the
factory]
Warning
1)
Diagnostic behavior can be changed.
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
948 Oscillation damping too high
Check process conditions
0x8000168
Status signal
S
Diagnostic behavior
Warning
12.9 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via local display → 219
• Via "FieldCare" operating tool → 222
• Via "DeviceCare" operating tool → 222
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
Previous diagnostics