WARNING/CAUTION/NOTE

Carefully read and strictly follow all Warnings and Cautions listed in this user guide. Servicing and maintenance should 
only be done by qualified personnel.

WARNING

 

9

Always hold the Bracket with both hands when you manipulate it. 

 

9

Operate the bracket only as described in this user guide.

 

9

Do not modify the Bracket, or any components of Technimount systems. Modifying the product can cause 

unpredictable operation resulting in injury to the patient or operator. 

 

9

It is the responsibility of the operator of the Bracket or any of its options, to ensure equipment being used with the 

Technimount products meets the installation specifications. Injury may result if  non-compatible products are used 

with Technimount System products.

 

9

Do not attempt to operate, install or remove the equipment while mobile. Ensure that installation and removal of 

equipment is done while vehicle or aircraft is immobilized.

 

9

Verify that the bracket is inserted correctly into a mounting system and locked with whichever disc is used (standard 

bottom disc or back anti-rotation disc). If equipment is improperly installed and locked, it can cause the equipment 

to fall and cause injury to the patient or operator and/or damage to the medical equipment.

 

9

Never use the Bracket if not properly installed and locked on a Surface Base, Safety Arm System on a cot/stretcher, 

on a Wall Mount or Floor Mount. If not installed properly, the medical device could fall during transport and may 

cause injury to the patient or operator and/or damage to the product.

 

9

Do not allow untrained staff to assist in the operation of the mounting system. Untrained technicians/staff can 

cause injury to the patient or themselves.

 

9

Do not place additional equipment or items onto the Bracket or mounts other than approved Technimount Systems 

or components, as this could damage the product and may result in injury to the patient or operator.

 

9

Operators must be able to able to safely lift the total weight of the Bracket System and Medical Device.

 

9

Do not attach restraints to the Bracket frame. It may result in damage to the System and/or result in injury to the 

patient or operator.

 

9

Always reposition the equipment with care, avoid any improper pushing or violent manipulation, as it may result in 

unbalanced equipment and may cause injury to the patient or operator and/or damage to the product.

 

9

Improper maintenance can cause injury to patient or operator or damage to the product. Maintain the Bracket 

and Mounting Systems as described in this user guide. Use only Technimount approved parts and maintenance 

procedures. Using unapproved parts and procedures could cause unpredictable operation and/or injury and will 

void the product warranty.

 

9

Failure to properly clean or dispose of corrosive products will increase the risk of premature damage and may 

cause injury to the patient or operator.

 

9

Any installation on a surface, wall or on the structure of the vehicle or the aircraft has to be validated or inspected 

in order to ensure it maintains the integrity of the vehicle or the aircraft.

Summary of Safety Precautions

Содержание BRACKET PRO 75 Series

Страница 1: ...BRACKET PRO SERIE 75 USER GUIDE EMS TM ...

Страница 2: ... or translated into any language in any form or by any means without the written permission of Technimount System or its affiliate companies Trademark Attributions Bracket Pro Serie Technimount EMS are registered trademarks of Technologies CGC Inc Tempus LS is a registered trademark from Philips In order to ensure continuous improvements to our products and services we invite you to communicate yo...

Страница 3: ...339 6351 NOTE For any issues with your Technimount product its components or for any technical questions during the installation or maintenance processes please send an email to techsupport technimount com Please have the serial number of your Technimount product available as shown in the figure below when calling Technimount Technical Support Include the serial number in all written communication...

Страница 4: ...eral Operating Guidelines for the Safety MD Transporter 22 Operating the Bracket on the Safety MD Transporter 23 Installing the Bracket onto the Safety MD Transporter 23 Rotating the Bracket on the Safety MD Transporter 24 Removing the Bracket with the Medical Device from the Safety MD Transporter 25 General Operating Guidelines for the Safety Arm System 26 Operating the Bracket on the Safety Arm ...

Страница 5: ...IMITED RESPONSIBILITY AND WARRANTY 36 INTERNATIONAL WARRANTY CLAUSE 36 Return Policy 37 Prior to 30 Days 37 Prior to 45 Days 37 Prior to 60 Days 37 Return Authorization 37 Damaged Merchandise 38 Claim Process 38 Required Information 38 Findings and Conclusion 38 questions about our policy 38 SAE CERTIFICATION 39 ...

Страница 6: ...ty Arm System on a Standard Surface Base on a Safety MD Transporter or with Technimount s floor mount or wall mount systems These solutions provide a higher flexibility for different EMS protocols and applications The Bracket Pro Serie 75 is compatible with the floor platform of air ambulances in Europe It can also be used with Technimount s wall mount systems This mounting system is compatible wi...

Страница 7: ...BP75 UG 202110 01 7 Return To Table of Contents Introduction POSSIBLE MOUNTING OPTIONS FOR THE BRACKET PRO SERIE 75 Safety MD Transporter Wall Mount System Ground Ambulance in Europe Safety Arm System Standard Surface Base ...

Страница 8: ...75 UG 202110 01 www technimount com Introduction SYMBOLS AND DEFINITIONS SYMBOL DESCRIPTION Warning and Caution special attention is required Consult accompanying documents Safe working load symbol and Load balance symbol Pinch Point ...

Страница 9: ...serious injury It may also describe potential serious adverse reactions and safety hazards Caution Alerts the reader of a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property This includes special care necessary for the safe and effective use of the device to avoid damage that may occur as a ...

Страница 10: ...ront View Introduction Rubber pads to protect the medical device from scratches Standard Bottom Disc compatible with all of Technimount s mounting solutions Main assembly is attached to the Tempus LS with a quick release and locking mechanism Acetal feet to protect the disc and Tempus LS when placed on the ground ...

Страница 11: ...LUSTRATION CONTINUED BRACKET PRO SERIE 75 CONTINUED Back View Introduction Anti Rotation Micro Disc is used for vertical support on different wall or arm systems Technimount molded polyurethane transport handle Bracket is made from high density aluminum with black anodized finish ...

Страница 12: ... Discs Standard Bottom Disc Anti Rotation Micro Disc Transportation Compatibility Systems Compatible with Technimount s Standard Surface Base Safety Arm System for Stryker Cots PowerPRO XT and MX Pro Floor Mount Platform Flight Wall Mount System Safety MD Transporter Features Access to all Connectors and Accessories No Screws Attached to the Medical Device Red Quick Release to Easily Insert and Re...

Страница 13: ...tomerservice technimount com Technimount E M S Holding Inc C O Regulatory Affairs 3770 Jean Marchand Street Suite 100 C Quebec QC G2C 1Y6 Canada T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com WARNING The end user and administrator are responsible to validate regulations and standards for safety in their region in order to comply with applicable safety regulations Technimoun...

Страница 14: ...operly the medical device could fall during transport and may cause injury to the patient or operator and or damage to the product 9 Do not allow untrained staff to assist in the operation of the mounting system Untrained technicians staff can cause injury to the patient or themselves 9 Do not place additional equipment or items onto the Bracket or mounts other than approved Technimount Systems or...

Страница 15: ...ausing premature corrosion of critical components 9 When ordering it is important to identify on which mounting system you need to mount the bracket on as there are different systems for different EMS applications e g EMS ground or air mounting systems 9 A maintenance program should be established for all Technimount equipment Preventive maintenance may need to be performed more frequently based o...

Страница 16: ... must work properly before being put into service 5 Refer to the identification of components at the beginning of this user guide if needed TOOLS REQUIRED FOR INSTALLATION Below are the basic tools required for installation If you have any questions regarding the installation or need assistance please contact our Technical Support team at techsupport technimount com 9 Philips Screwdriver NOTE 9 El...

Страница 17: ... be installed on the Tempus LS monitor from Philips 2 To install the Tempus LS monitor pull the red handle and raise the top part until it locks into the top position 3 Align the Tempus LS in front of the bracket while it is still raised Installation Guide INSTALLATION OF THE BRACKET ONTO THE TEMPUS LS MONITOR ...

Страница 18: ...tents INSTALLATION OF THE BRACKET ONTO THE TEMPUS LS MONITOR CONTINUED 4 Install the Tempus LS into the Main Assembly 5 Pull the red handle and lower the top part until it locks into the bottom platform 6 TheTempusLSmonitorisreadyforEMStransport Installation Guide ...

Страница 19: ... on the Philips Tempus LS monitor 2 To install the Tempus LS Smart Mount monitor pull the Red Quick Release Mechanism and raise the top part until it locks into the top position 3 Align the Smart Mount in front of the bracket while it is still raised Installation Guide INSTALLATION OF THE BRACKET ONTO TEMPUS LS MONITOR WITH SMART MOUNT ...

Страница 20: ...NTO THE TEMPUS LS MONITOR WITH SMART MOUNT CONTINUED 4 Install the Smart Mount into the Main Assembly 5 Align screw holes 6 Install the four Stainless Steel Flat Head Philips Screws 10 32 X 9 16 in 7 Tighten all four screws 8 Install the Tempus LS monitor onto the Smart Mount as described in its user manual Installation Guide ...

Страница 21: ...Contents Return To Table of Contents INSTALLATION OF THE BRACKET ONTO THE TEMPUS LS MONITOR WITH SMART MOUNT CONTINUED Installation Guide 9 Lower the top part until it locks into the bottom position 10 The Tempus LS Smart Mount monitor is ready for EMS transport ...

Страница 22: ...ES FOR THE SAFETY MD TRANSPORTER 9 Use the Safety MD Transporter only as described in the User Guide 9 Read all the labels and instructions on the product before using the Safety MD Transporter 9 Always inspect the Micro Base and the Safety MD Transporter for any issue or damage before every use 9 Installing or removing the Safety MD Transporter requires a minimum of one 1 trained operator 9 Do no...

Страница 23: ... on the back can be inserted into the Micro Base located on the Safety MD Transporter 2 Ensure that your bracket has an Anti Rotation Micro Disc on the back to install the medical device onto the Safety MD Transporter 3 While using both hands slowly insert Anti Rotation Micro Disc located on the back of the bracket into the Micro Base WARNING 9 Ensure that the bracket is securely locked into the M...

Страница 24: ...RANSPORTER CONTINUED INSTALLING THE BRACKET ONTO THE SAFETY MD TRANSPORTER CONTINUED Operation Guide 4 A Bracket Pro Serie 75 installed on the Safety MD Transporter ROTATING THE BRACKET ON THE SAFETY MD TRANSPORTER 1 To rotate the medical device 360 turn the Safety MD Transporter clockwise or counter clockwise ...

Страница 25: ...ONTINUED REMOVING THE BRACKET WITH THE MEDICAL DEVICE FROM THE SAFETY MD TRANSPORTER 1 To remove the bracket from the Safety MD Transporter press the Red Quick Release Mechanism located on the back Micro Base 2 While holding the Red Quick Release Mechanism lift the bracket upwards until removed and release the Red Quick Release Mechanism ...

Страница 26: ...m of one 1 trained operator 9 Do not remove the Safety Arm System or device from the system while moving with the cot stretcher 9 Stay with the patient and control the Safety Arm System and device during transport at all times GENERAL OPERATING GUIDELINES FOR THE SAFETY ARM SYSTEM Operation Guide WARNING 9 Never let a patient grab manipulate or hold the Safety Arm System the Standard Surface Base ...

Страница 27: ... is installed Wall Mount Floor Mount Safety Arm System etc 3 To insert the bracket into the Standard Surface Base slide into the base horizontally from the front of the base 4 Ensure the bracket is locked securely in the Standard Surface Base before manipulating the device or transporting the stretcher ROTATING THE BRACKET ON THE STANDARD SURFACE BASE 1 To rotate the medical device 360 turn the br...

Страница 28: ...ROM THE STANDARD SURFACE BASE 1 To remove the bracket from the Standard Surface Base press the Red Quick Release Mechanism located on the front of the Standard Surface Base 2 While holding the Red Quick Release Mechanism slide outwards horizontally to remove from the Standard Surface Base 3 Hold the Tempus LS from the middle of the unit to gently remove from the base ...

Страница 29: ...dels Available refer to Customer Service WARNING Do not modify any components of these system Modifying the products can cause unpredictable operation resulting in injury to the patient or operator NOTE Technimount reserves the right to change Part Numbers and Products without notice Please refer to Customer Service to ensure product part number availability For any replacement parts or for repair...

Страница 30: ...ower washer can accelerate removal of contaminants collected during the use of the product 9 Rinse with clean water 9 Towel dry all components 9 Allow all other components to air dry 9 Avoid over saturation and ensure that the product does not stay wet longer than the cleaner manufacturer s guidelines for proper disinfecting WARNING 9 When cleaning always use appropriate personal protection equipm...

Страница 31: ...ervice WARNING 9 Failure to properly clean contaminated parts or components will increase the risk of exposure to blood borne pathogens and may cause injury to the patient or the operator PREVENTIVE MAINTENANCE A preventive and regular maintenance program should be established for all Technimount equipment Preventive maintenance may be required more frequently based on the usage level of the produ...

Страница 32: ...for maintenance either in months M or hours H in accordance with your Service s current maintenance practices and protocols ITEM ROUTINE 1 M OR 2 H 3 M OR 6 H 6 M OR 12 H 12 M OR 24 H Bracket 9 Inspect Bracket Inspect the screws Inspect the Red Quick Release Mechansim Ensure it is working properly and there are no particles obstacles or debris Inspect back screws 9 Verify both discs Verify that th...

Страница 33: ...www technimount com BP75 UG 202110 01 33 Return To Table of Contents Maintenance Guide INSPECTION AND MAINTENANCE RECORD PREVENTIVE MAINTENANCE PERFORMED BY DATE TIME ...

Страница 34: ...n To Table of Contents 34 BP75 UG 202110 01 www technimount com Maintenance Guide TRAINING RECORD TRAINING DATE TRAINING METHOD USER GUIDE IN SERVICE IN CLASS ETC TRAINEE NAME BASIC TRAINING TRAINING UPDATE ...

Страница 35: ...ilver Anodized Locking Assembly Block Without Bag Kit Includes Protection Rubber Locking Assembly Block and 2 Phillips Stainless Steel Pan Head Screws with Nylon Patch 10 32 X 1 2 in TCM 044 014 Black Anodized Locking Assembly Block With Bag Kit Includes Protection Rubber Locking Assembly Block and 2 Phillips Stainless Steel Pan Head Screws with Nylon Patch 10 32 X 1 2 in TCM 044 019 Screw Kit for...

Страница 36: ...replaced after the Warranty Period has expired the Warranty Period for the repair or replacement will expire three 3 months after the date of repair or replacement LIMITED RESPONSIBILITY AND WARRANTY The Limited Warranty does not apply to normal wear that should result from normal use It does not apply when the Hardware or any component is opened disassembled or repaired by someone not authorized ...

Страница 37: ...YS Product must be unused undamaged and in its original packaging The product return request must be provided in writing and it must be approved by TECHNIMOUNT SYSTEM prior to returning the product Returns will not be approved on modified or damaged item Charges may apply if damaged or missing items Customer is responsible for a 25 restocking fee PRIOR TO 60 DAYS Product must be unused undamaged a...

Страница 38: ...e RMA form online at https www technimount com service support service support overview REQUIRED INFORMATION Serial number Invoice number Identification of the issue Upload the pictures of the issue Submit the form A confirmation of the receipt of the claim will be sent to you within 2 business days If you have not received a confirmation Email by then please contact techsupport technimount com If...

Страница 39: ...www technimount com BP75 UG 202110 01 39 Return To Table of Contents SAE Certification ...

Страница 40: ...e it matters most Technimount E M S Holding Inc 3770 Jean Marchand Street Suite 100 C Quebec QC G2C 1Y6 Canada T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com 2021 Technimount E M S Holding Inc All Rights Reserved Printed in Canada ...

Отзывы: