Maintenance Guide

CLEANING THE BRACKET SYSTEM

CLEANING PROCESS

WARNING

Follow  your  EMS  service's  protocol  or  medical  device  manufacturer's  user  guide  for  the  recommended  cleaning 

procedure of your medical devices. Cleaning procedure below does NOT provide recommendations for cleaning of the 

medical device. The following cleaning guidelines consider the Technimount mounting assemblies and components. 

Remove medical device from the bracket if thorough cleaning is needed.

The Bracket can be cleaned  by using a pressure washer. The bracket may show some signs of oxidation or discoloration 

from  continuous  washing,  however,  degradation  of  the  cot/stretcher’s  performance  or  functionality  will  not  be 

affected due to power washing if cleaning instructions are properly followed. 

The material used to manufacture the Bracket is high-grade aluminum, it is very durable and resistant  to corrosion, 

however high levels of  sodium or a high concentration  of  sodium or acid  may prematurely corrode the product. 

Always rinse with soft soap and water if exposed.

 

9

Use of a power washer can accelerate removal of contaminants collected during the use of the product.

 

9

Rinse with clean water.  

 

9

Towel dry all components.

 

9

Allow all other components to air dry. 

 

9

Avoid over saturation and ensure that the product does not stay wet longer than the cleaner manufacturer’s 

guidelines for proper disinfecting.

WARNING

 

9

When cleaning, always use  appropriate personal protection equipment (PPE) based on established protocols (e.g., 

gloves, eyewear, etc.).

CAUTION

 

9

Do not steam clean or use ultrasonic cleaners on the system or any of its components.

 

9

Do not immerse the metal parts/components in water.  

 

9

Maximum water temperature should not exceed 180°F/82°C.

 

9

Maximum water pressure should not exceed 1500 psi/103.5 BAR . If a pressure washer is being used to clean the 

unit, the pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.

CLEANING SOLUTIONS

Phenolic type or quaternary type disinfectants can be used (excluding Virex® TB). Iodophor type disinfectants, Peracetic 

Acid, Chlorine, and especially Hydrogen Peroxide are not recommended for use because staining and corrosion may 

result. 

 

Suggested cleaners for the Bracket:

 

9

Quaternary Cleaners (active ingredient - ammonium chloride)

 

9

Phenolic Cleaners (active ingredient - o-phenyl phenol)

 

9

Chlorinated Bleach Solution (3.25% - less than 1-part bleach to 100 parts water)

Содержание BRACKET PRO 121-FL Series

Страница 1: ...BRACKET PRO SERIE 121 FL USER GUIDE EMS TM ...

Страница 2: ...slated into any language in any form or by any means without the written permission of Technimount or its affiliate companies Trademark Attributions Bracket Pro Serie Technimount EMS are registered trade marks of Technologies CGC Inc Spectrum IQ is a registered trademark from Baxter In order to ensure continuous improvements to our products and services we invite you to communicate your comments t...

Страница 3: ... 6351 NOTE For any issues with your Technimount product its components or for any technical questions during the installation or maintenance please send an Email to techsupport technimount com Please have the serial number of your Technimount product available as shown in the figure below when calling Technimount Customer Service or Technical Support Include the serial number in all written commun...

Страница 4: ...6 Removing the Packaging for Installation 16 Installation of the Baxter Pumps onto the Bracket 17 PART NUMBERS 20 Kits and Other Options 20 MAINTENANCE GUIDE 21 Cleaning the Bracket System 21 Cleaning Process 21 Cleaning Solutions 21 Removal of Iodine Compounds 22 Preventive Maintenance 22 Lubrication 22 Inspection Process and Schedule 23 Maintenance Program 23 Inspection and Maintenance Record 24...

Страница 5: ...www technimount com BP121 FL UG 202111 01 5 Table of Contents Required Information 29 Findings and Conclusion 29 Questions about our policy 29 SAE CERTIFICATION 30 ...

Страница 6: ...em on a Universal Standard Base on a Safety MD Transporter or with Technimount s Floor Mount or Wall Mount Systems These solutions provide a higher flexibility for different EMS CCT protocols and applications The Bracket Pro Serie 121 FL is compatible with the floor platform of air ambulances in Europe It can also be used with the Wall Mount System for ground ambulances in Europe and North America...

Страница 7: ...202111 01 7 Return To Table of Contents Introduction POSSIBLE MOUNTING OPTIONS FOR THE BRACKET PRO SERIE 121 FL Safety MD Transporter Wall Mount System EU Proplate with 2 Micro Bases for IT Stretchers Safety MD Transporter Horizonatal ...

Страница 8: ...1 FL UG 202111 01 www technimount com Introduction SYMBOLS AND DEFINITIONS SYMBOL DESCRIPTION Warning and Caution special attention is required Consult accompanying documents Safe working load symbol and Load balance symbol Pinch Point ...

Страница 9: ...r serious injury It may also describe potential serious adverse reactions and safety hazards Caution Alerts the reader of a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property This includes special care necessary for the safe and effective use of the device to avoid damage that may occur as ...

Страница 10: ...Front View Introduction Integrated feet to protect the disc when placed on the ground Bottom Micro Disc compatible with Technimount s mounting soltuions Rubber pads to protect the medical device Stainless Steel IV pole hooks Technimount Molded Polyurethane Transport Handle High density aluminum with blue anodized finish ...

Страница 11: ...FL CONTINUED Back View Introduction IV Pole made of high density aluminum with silver anodized finish Anti Rotation Micro Disc is used for vertical support on different wall or arm systems Quick Release Mechanism to install the pumps Quick Release Knob to lower or raise the IV pole IV pole holder to raise or lower when needed ...

Страница 12: ...ertification Designed to Comply with SAE J3043 26 G FAA 14 CFR 23 applicable section Installation Device Retained on Main Assembly with Quick Release Mechansim Attachment Options 2 Locking Discs Bottom Micro Disc Anti Rotation Micro Disc Transportation Compatibility System Compatible with Technimount s Wall Mount Systems Safety MD Transporters Proplate with Micro Bases for Incubators Features Robu...

Страница 13: ...com for more information Technimount E M S Holding Inc C O Regulatory Affairs 3770 Jean Marchand Street Suite 100 C Quebec QC G2C 1Y6 Canada T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com WARNING The end user and administrator are responsible to validate regulations and standards for safety in their region in order to comply with applicable safety regulations Technimount is...

Страница 14: ... the mounting system Untrained technicians staff can cause injury to the patient or themselves 9 Do not place additional equipment or items onto the Bracket or mounts other than approved Technimount Systems or components as this could damage the product and may result in injury to the patient or operator 9 Operators must be able to able to safely lift the total weight of the Bracket System and Med...

Страница 15: ...ace possibly causing premature corrosion of critical components 9 When ordering it is important to identify on which mounting system you need to mount the bracket on as their are different systems for different EMS applications e g EMS ground mounting systems 9 A maintenance program should be established for all Technimount equipment Preventive maintenance may need to be performed more frequently ...

Страница 16: ...nstallation 3 Identify all of the product components and hardware prior to starting the installation 4 The Bracket System must work properly before being put into service 5 Refer to the identification of components at the beginning of this user guide if needed TOOLS REQUIRED FOR INSTALLATION No tools are required for installation If you have any questions regarding the installation or need assista...

Страница 17: ...nts 1 Place the bracket on a stable surface for installation 2 Before inserting the device locate the locking block on the medical device 3 Align the locking block with the support system on the bracket Installation Guide INSTALLATION OF THE BAXTER PUMPS ONTO THE BRACKET ...

Страница 18: ...n the Quick Release Mechansim and insert the medical device into the bracket 5 Proceed to install all pumps while following steps 1 to 4 6 Once properly locked the Baxter pumps are ready for EMS transport Installation Guide INSTALLATION OF THE BAXTER PUMPS ONTO THE BRACKET CONTINUED ...

Страница 19: ... the IV pole from the top of the IV pole holder 2 Slide in until you reach one of the grooves on the IV pole to hold into position 3 The IV pole is now ready to be used with IV bags WARNING 9 Do not use the IV pole during EMS transport Installation Guide INSTALLATION OF THE IV POLE ...

Страница 20: ...ht configuration for your stretcher WARNING Do not modify any components of these system Modifying the products can cause unpredictable operation resulting in injury to the patient or operator NOTE Technimount reserves the right to change Part Numbers and Products without notice Please refer to Customer Service to ensure product part number availability For any replacement parts or for repair plea...

Страница 21: ...a power washer can accelerate removal of contaminants collected during the use of the product 9 Rinse with clean water 9 Towel dry all components 9 Allow all other components to air dry 9 Avoid over saturation and ensure that the product does not stay wet longer than the cleaner manufacturer s guidelines for proper disinfecting WARNING 9 When cleaning always use appropriate personal protection equ...

Страница 22: ...ervice WARNING 9 Failure to properly clean contaminated parts or components will increase the risk of exposure to blood borne pathogens and may cause injury to the patient or the operator PREVENTIVE MAINTENANCE A preventive and regular maintenance program should be established for all Technimount equipment Preventive maintenance may be required more frequently based on the usage level of the produ...

Страница 23: ...t com Check each routine and replace damaged or worn out parts if necessary Follow timetable below for maintenance either in months M or hours H in accordance with your Service s current maintenance practices and protocols ITEM ROUTINE 1 M OR 2 H 3 M OR 6 H 6 M OR 12 H 12 M OR 24 H Bracket 9 Inspect Bracket Inspect the screws Ensure it is working properly and there are no particles obstacles or de...

Страница 24: ...Return To Table of Contents 24 BP121 FL UG 202111 01 www technimount com Maintenance Guide INSPECTION AND MAINTENANCE RECORD PREVENTIVE MAINTENANCE PERFORMED BY DATE TIME ...

Страница 25: ...hnimount com BP121 FL UG 202111 01 25 Return To Table of Contents Maintenance Guide TRAINING RECORD TRAINING DATE TRAINING METHOD USER GUIDE IN SERVICE IN CLASS ETC TRAINEE NAME BASIC TRAINING TRAINING UPDATE ...

Страница 26: ...l currently available for purchase Some of the parts identified on the assembly drawing parts in this user guide may not be individually available for purchase Please call Technimount Customer Service 1 888 639 2758 or at customerservice technimount com for availability and pricing DESCRIPTION PART NUMBER IV Pole TCM 052 037 Acetal Feet TCM 032 082 ...

Страница 27: ...eplaced after the Warranty Period has expired the Warranty Period for the repair or replacement will expire three 3 months after the date of repair or replacement LIMITED RESPONSIBILITY AND WARRANTY The Limited Warranty does not apply to normal wear that should result from normal use It does not apply when the Hardware or any component is opened disassembled or repaired by someone not authorized b...

Страница 28: ...45 DAYS Product must be unused undamaged and in its original packaging The product return request must be provided in writing and it must be approved by TECHNIMOUNT prior to returning the product Returns will not be approved on modified or damaged item Charges may apply if damaged or missing items Customer is responsible for a 25 restocking fee PRIOR TO 60 DAYS Product must be unused undamaged and...

Страница 29: ...the RMA form online at https www technimount com service support service support overview REQUIRED INFORMATION Serial number Invoice number Identification of the issue Upload the pictures of the issue Submit the form A confirmation of the receipt of the claim will be sent to you within 2 business days If you have not received a confirmation Email by then please contact techsupport technimount com ...

Страница 30: ...Return To Table of Contents 30 BP121 FL UG 202111 01 www technimount com SAE Certification ...

Страница 31: ...www technimount com BP121 FL UG 202111 01 31 Return To Table of Contents ...

Страница 32: ...e it matters most Technimount E M S Holding Inc 3770 Jean Marchand Street Suite 100 C Quebec QC G2C 1Y6 Canada T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com 2021 Technimount E M S Holding Inc All Rights Reserved Printed in Canada ...

Отзывы: