Specification
Figures quoted are at the base of the well at the time of calibration.
Temperature range:
5°C/9°F above ambient to 430°C/806°F
Over-temperature limit:
450°C/842°F (approximately)
Display resolution:
0.1°
200 degrees C
400 degrees C
Accuracy ±0.3°C
±0.5°C
Stability (after 15 minutes)
±0.03°C
±0.05°C
Well to well uniformity
0.015°C
0.025°C
Heat up time 20° C to 400°C:
12 minutes
Cool down 400°C to 100°C:
21 minutes
Immersion Depth:
4.5" (114.3mm)
Fan
Cooling:
Automatic
Weight:
9
lbs
(4
Kg)
Dimensions* (H x W x D):
8.75 x 8 x 8 inches/222.25 x 203.2 x 203.2 mm
*excluding the carrying strap
Electrical supply
Voltage Cycles
Power
230V 50/60Hz 900W
120V 50/60Hz 900W
Note:
The above specifications are quoted for an ambient temperature range of 10°C/50°F to 30°C/86°F.
Outside this range, the quoted figures may deteriorate but the unit will still work safely.
-2-