P A G E  12

R e f r i g e r a t e d   B a t h s   U S E R ’ S   M A N U A L

9  Fill the bath to between the minimum and maximum levels 

which are 65mm and 20mm from the top of the bath.

10 Fridge baths are supplied with a lid and this should be used 

whenever possible..

Cleaning your unit

Before cleaning your unit, disconnect from the power supply and 
allow to cool to ambient temperature.

You can clean the case of the refrigerated unit with a cloth dipped 
in water or ethanol (methanol or formaldehyde can also be used). 
No part of the case or cover should be immersed in the solvents.

Do not use acetone or abrasive cleaners.

Packing

When you receive your new refrgerated bath, remove all the 
packing.  check the unit for any transit damage.  If there is any 
transit damage at all inform Techne Sales Department or your 
Techne dealer immediately.
When unpacking the unit, check that the following have been 
removed from the packing: a power cable (RB-5A only); a 
decontamination certificate and a Guarantee card.
check that the unit is the correct rating for your power supply by 
looking on the Rating Plate which is on the rear of the unit.
Within the guarantee period, shown on the Guarantee card, we 
undertake to supply replacements free of charge for parts which 
may, on examination, prove to be defective, provided that the 
defect is not the result of misuse, accident or negligence.  Any 
instrument requiring service under this guarantee should be sent 
to the supplier through whom it was purchased, or, in the case of 
difficulty, it should be carefully packed in its original packing and 
consigned, carriage paid, to us.  Techne takes no responsibility for 
returned goods damaged in transit.

Returned goods will not be processed without a Returns 

Authorisation number. Call + (44) (0)1785 810475 for a 

number. 

On all correspondence, please quote the Serial Number 

in full and/or the Sales Order Number.

  Please write the Returns 

number on the outside of any packing.

 

oPeRAtIon

Содержание RB-12A

Страница 1: ...P A G E 1 R e f r i g e r a t e d B a t h s U S E R S M A N U A L RB 5A RB 12A RB 22A Refrigerated Baths OPERATOR S MANUAL Issue 12 Date of issue 22nd March 2011 Techne is a trademark Techne 2012...

Страница 2: ......

Страница 3: ...ais 5 Deutsch 6 Espa ol 7 Introduction 8 Operator Safety 8 Specification 9 Siting Considerations 10 Setting up the unit 10 Cleaning your unit 12 Packing 12 Operation 13 Precautionary Measures 13 Oper...

Страница 4: ...ing meanings I mains switch On O mains switch Off After use When you have finished heating samples remember that partsoftheunit thetubes blocksandassociatedaccessories may be very hot Take the precaut...

Страница 5: ...reil sur un plan de travail ou surface plane ou le cas ch ant dans une hotte d aspiration en s assurant que les trous d a ration situ s sous l appareil ne sont pas obstru s 5 Les symboles situ s sur o...

Страница 6: ...e Arbeitsfl che bzw falls erforderlich unter einen Laborabzug Beachten Sie da die Entl ftungsrippen an der Ger teunterseite immer frei zug nglich sind 5 Die Symbole auf oder neben dem EIN AUS Schalter...

Страница 7: ...elasentradasdeaireenlaparteinferiornoqueden obstruidas 5 Loss mbolosqueseencuentranenocercadelinterruptor de alimentaci n tienen los siguientes significados I Interruptorprincipalencendido O Interrupt...

Страница 8: ...lator TE 10 series or TU 20 series Each unit fridge or controller has its particular application which should be considered when setting up a temperature controlled system The refrigerated baths have...

Страница 9: ...lation system condition of the working liquid heat input and so on RB 5A RB 12A RB 22A Minimum temperature 20 o C 35 C 30 C Cooling capacity at 20 o C 145 W 240W 240W 0 o C 145 W 240W 240W 10 o C 110...

Страница 10: ...at of the mains supply Connect the mains cable to a suitable plug as in the following table Note that the unit must be earthed to ensure proper electrical safety Connections 230V 110 120V Live brown b...

Страница 11: ...hich will cover the full range of working temperatures The RB 12A RB 22A units are designed to work up to 40o C with occasional peaks to 100o C The RB 5A can be used up to 150 C but unit should be tur...

Страница 12: ...following have been removed from the packing a power cable RB 5A only a decontamination certificate and a Guarantee Card Check that the unit is the correct rating for your power supply by looking on t...

Страница 13: ...temperatures below 40o C 5 IN NO CIRCUMSTANCES MAY A REFRIGERATED UNIT BE SUBJECTED TO BATH LIQUID TEMPERATURES ABOVE 100o C Operating your unit 1 Switch on the thermoregulator as instructed in the a...

Страница 14: ...ter or by balls from a ball blanket or in any other way Also check that the pipe system if you are using the fridge unit for an external system is not kinked or restricted in any way 4 If the liquid o...

Страница 15: ...as to the gas charge for each of the units Unit Refrigerant Charge RB 5A R134a 85 gm RB 12A R404a 250 gm RB 22A R404a 250 gm There have been several improvements in design which have varied the charg...

Страница 16: ...ge to units which are not properly packed for shipping if in doubt contact your supplier Fuses If the power light on the power switch is lit the fuse may have blown Make sure there is no external caus...

Страница 17: ...P A G E 17 R e f r i g e r a t e d B a t h s U S E R S M A N U A L...

Страница 18: ...P A G E 18 R e f r i g e r a t e d B a t h s U S E R S M A N U A L...

Страница 19: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: