![Techne IFB-51 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/techne/ifb-51/ifb-51_instruction-manual_1067156018.webp)
DESCRIPTION OF COMPONENTS AND FRONT PANEL
CONTROLS
Figure 1: Components and controls – IFB-51 shown
Item
Description
Function
1
Fluidised bath chamber
Heated chamber containing the fluidised aluminium oxide.
2
Operating instructions
Quick operating guide.
3
Eurotherm 3216
temperature controller
Used to set the required bath set point temperature. The
control parameters have been factory pre-set.
4
Power indicator (green)
Lights green when there is mains power to the unit.
5
Heater indicator (amber)
Lights amber when the heaters are on. Will flash on and off
when the bath is at set temperature.
6
Air flow table
Indicates approximate air flow according to temperature.
7
Flow meter
A float indicates the fluidisation air pressure set by the control
valve. The flow rate indicated by the float in the tube is on a
scale of 0.5 to 4.0 CFM (cubic feet per minute).
8
Air flow valve
Controls the amount of air for fluidising the bed.
9
Fuses
The unit is fitted with 2 x 25A (IFB-51) or 2 x 30A (IFB-52) fuses
10
Rating plate
Details the model number, serial number and power rating.
11
Mains power cord
Connect to mains power supply.
12
Compressed air inlet
The air supply should be connected via suitable flexible hose.
18