operatIon
preparation
1. Place the aluminium insert blocks to be used into the Dri-Block
®
unit. The blocks should touch each
other and be positioned centrally.
2. Loosen the locking knob on the carriage. Remove the gas chamber, disconnecting the gas pipe if
necessary, and lay the chamber upside down on a suitable work surface.
3. Insert the hyperdermic needles in the pattern to suit the test tubes to be placed in the heater.
The needle guide holes are labelled with a letter corresponding with the type of insert block used in the
heater. These are listed in the following table. As an example if you use the block for twelve 15mm test
tubes (e.g. block number F3506) you would place a needle in all the holes labelled C.
Only Sample Concentrators with the part number FSC496D or a replacement head part number 6106467
will have the 96 well array.
Tube size
Hole array
Hole label
& diameter
5 x 6 deep
30
A
4 x 5 deep
20
B
3 x 4 deep
12
C
8 hole off-set array, e.g. F4490
D
6