Techne FSC400D Скачать руководство пользователя страница 8

operatIon

preparation

1.  Place the aluminium insert blocks to be used into the Dri-Block

®

 unit. The blocks should touch each 

other and be positioned centrally.

2.  Loosen the locking knob on the carriage. Remove the gas chamber, disconnecting the gas pipe if 

necessary, and lay the chamber upside down on a suitable work surface.

3.  Insert the hyperdermic needles in the pattern to suit the test tubes to be placed in the heater.

The needle guide holes are labelled with a letter corresponding with the type of insert block used in the 
heater. These are listed in the following table. As an example if you use the block for twelve 15mm test 
tubes (e.g. block number F3506) you would place a needle in all the holes labelled C.

Only Sample Concentrators with the part number FSC496D or a replacement head part number 6106467 
will have the 96 well array.

 

Tube size    

 

Hole array    

      Hole label

 

& diameter  

 

 

 

 

5  x  6 deep 

30 

A

 

4  x  5 deep 

20 

B

 

3  x  4 deep 

12 

C

 

8 hole off-set array, e.g. F4490   

D

6

Содержание FSC400D

Страница 1: ...Sample Concentrator operator s manual Issue 14 10 15...

Страница 2: ......

Страница 3: ...fety 3 Specification 3 Installation 3 Diagram of Sample Concentrator 4 Guarantee 5 operation 6 Preparation 7 Operation 8 ADDITIONAL INFORMATION 9 User maintenance 9 Solvent evaporation temperatures 10...

Страница 4: ...rected over the surface of the samples displaces the evaporated solvent from above the liquid s surface This significantly increases the rate of sample concentration The unit provides accurate and rep...

Страница 5: ...to suit Techne Dri Blocks Gas Intake Nozzle Diameter 6 35 mm Height 530 mm Width 240 mm Depth 295 mm Weight 3 5 Kg Installation For packing purposes the Sample Concentrator is supplied unassembled in...

Страница 6: ...xcess of 2 psi will cause gas to be vented and wasted 4 Place the Techne Dri Block heater onto the base of the Sample Concentrator It is positioned by means of two spigots on the base board which alig...

Страница 7: ...under this guarantee should be obtained from the supplier Notwithstanding the description and specification s of the units contained in the user manual Techne hereby reserves the right to make such ch...

Страница 8: ...test tubes to be placed in the heater The needle guide holes are labelled with a letter corresponding with the type of insert block used in the heater These are listed in the following table As an ex...

Страница 9: ...ion Build up the desired pattern of needles to suit the arrangement of test tubes to be used Do not put a needle in any position which will not have a corresponding sample to be concentrated as this w...

Страница 10: ...required height above the liquid surface 3 Set the heater to the required temperature and adjust the gas flow to the desired rate Remember not to exceed 2 psi 4 To increase the rate of concentration t...

Страница 11: ...ad After many repeated needle insertions the self sealing property of the sealing pad may start to degrade Spare pads can be obtained from your supplier or from Techne Replacing the pad is simply acco...

Страница 12: ...35 Atmospheric 1 2 dichloroethane Ethylene chloride 83 210 1 2 dichloroethene Cis acetylene dichloride 60 479 1 2 dichloroethene Trans acetylene dichloride 48 751 2 propan 2 yloxypropane Isopropyl eth...

Страница 13: ...ethane Carbon tetrachloride 76 300 1 1 2 2 tetrachloroethene Tetrachloroethylene 121 53 Oxolane Tetrahydrofuran THF 67 357 Toluene Methylbenzene 111 77 1 1 2 trichloroethene Trichloroethylene 87 183 O...

Страница 14: ...B B B B B B B B B B B B B B B A A A A A A A A A A A A A A A A A A A A A A A A A A D D D D D D A A A A A A A A A A A A A A A C C C C C C C C C C C C C C C C C F E E E E E E F F F F F B B B B B B B B B...

Страница 15: ...7210 Pack of 100 x 127mm long needles FSC4NCS Pack of PTFE coated needles 100 x 76mm long FSC4NCL Pack of PTFE coated needles 100 x 127mm long 6101608 Spigot 6101604 Sealing pad 6101605 Support strip...

Страница 16: ...l contact bibby scientific com Web www bibby scientific fr Bibby Scientific US 3 Terri Lane Suite 10 Burlington NJ 08016 USA Tel 1 609 589 2560 Fax 1 609 589 2571 Email labproducts techneusa com Web w...

Отзывы: