![Techcomp UV 2600 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/techcomp/uv-2600/uv-2600_instruction-manual_1066383047.webp)
CHAPTER 4 MAINTENANCE
44
Figure 4-18 Replacement and adjustment of D2 Lamp
Figure 4-19 Relative position of D2 lamp facula and light-entrance slit
Confirm
D2
lamp facula at
this position.
Viewfinder
Screw
B
Nut B
D2 lamp assembly
D2 lamp facula
Light-entrance
slit
Normal status
Horizontal deviation
Vertical deviation