3. Installation
3.4 Switching the Instrument ON
Caution
Before the instrument is switched on for the first time after installation,
it should be left to stand for at least 3 hours, so there is no possibility of
condensation causing a short circuit.
•
Ensure the computer is switched OFF and the instrument's mains power
switch in the back panel of the instrument is in the OFF position.
•
Connect the computer to the instrument with the delivered RS 232 interface
cable.
•
Insert the power cable into the mains power socket (with protective earth
connection) in the back panel of the instrument.
•
Connect the connecting cable to the computer and the instrument. All
connected devices must be proved and listed with regulations EN 60950,
UL 1950 or CSA C22.2 No. 950 for Data Processing Devices
•
Switch the instrument ON using the switch in the back panel of the
instrument.
Caution
When installing or uninstalling the instrument ensure that the instrument
and the computer are both switched off and disconnected from the mains
before the RS 232 interface cable is connected or removed.
Rear View
On/Off Switch
0 = OFF
1 = ON
RS 232
The instrument is now ready to be used with a suitable software program.
If the instructions given in this manual are not correctly performed the
instrument may be damaged or procedures may not be correctly performed.
Consequently the safety of the instrument can not be guaranteed.
To operate the software, refer to the XFLUOR manual.